Key result
AViTA BPM636 upper-arm monitor passes universal validation criteria within ~1 mmHg of reference BP.
Why the study?
The clinical accuracy of the AViTA BPM636 upper-arm blood pressure monitor in a combined pediatric and adult population according to the ANSI/AAMI/ISO Universal Standard was evaluated.
Does the AViTA BPM636 oscillometric upper-arm blood pressure monitor provide accurate blood pressure measurements compared to reference standards in a combined pediatric and adult population?
Cross-Sectional (n=85)
Does the AViTA BPM636 oscillometric upper-arm blood pressure monitor provide accurate blood pressure measurements compared to reference standards in a combined pediatric and adult population?
The AViTA BPM636 upper-arm blood pressure monitor passed ANSI/AAMI/ISO validation criteria and is clinically accurate for home use in both children and adults.
May support home BP monitoring across ages with validated devices; extends options meeting updated universal accuracy standards.
OBJECTIVE: To evaluate the clinical accuracy of the AViTA BPM636 oscillometric upper-arm blood pressure monitor in the tested pediatric and adult population, including children aged 5-12 years and adults aged 21-82 years with arm circumferences ranging from 15 to 42 cm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2018/Amd2:2024). METHODS: The AViTA BPM636 was evaluated in 85 participants (including 35 pediatric and 50 adult subjects) with upper-arm circumferences ranging from 15 to 42 cm, and two cuffs of different sizes (15 to 22 and 22 to 42 cm) were used in this study. The recruitment fulfilled the requirements of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) for a combined pediatric and adult population using the same arm sequential blood pressure measurement method. RESULTS: For criterion 1, the mean ± SD of the differences between the test device and reference blood pressure measurements was 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP. For criterion 2, the SD of individual mean differences for SBP (5.44 mmHg) and DBP (4.98 mmHg) remained well within the maximum permissible limits of 6.88 and 6.95 mmHg, respectively, thereby satisfying the validation criteria. CONCLUSION: The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.
No takes yet. Share an insight, caveat, or question.
Lin et al. (2026) reported a cross-sectional. AViTA BPM636 oscillometric upper-arm blood pressure monitor vs. Reference blood pressure measurements was evaluated on Differences between test device and reference blood pressure measurements (Criterion 1) (MD 0.93 ± 6.11 mmHg (SBP) and -0.20 ± 5.64 mmHg (DBP)). The AViTA BPM636 blood pressure monitor met validation criteria with mean differences of 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP compared to reference measurements.