Key result
Surgical removal via exploratory laparotomy and external iliac arteriotomy was required to retrieve a ruptured intra-aortic balloon pump after unsuccessful endovascular snaring.
Case Report (n=1)
This case demonstrates the management of a complex IABP rupture and entrapment requiring a hybrid surgical and endovascular approach.
Hybrid retrieval may be needed after failed endovascular snaring of ruptured IABP; this case leaves open optimal prevention and device strategies.
A 50-year-old woman required an emergent operative exploration after an unsuccessful attempt at bedside removal of a ruptured intra-aortic balloon pump (IABP). Initial fluoroscopy demonstrated that the IABP tip moved dynamically with blood flow, but retraction and advancement caused significant bowing of the proximal external iliac artery. After unsuccessful endovascular snaring and resheathing, an exploratory laparotomy was required to remove the entrapped IABP via a longitudinal external iliac arteriotomy. Proximal to the patch angioplasty, an occlusive dissection flap was identified, requiring stent placement. Back table examination showed an exposed metallic knob, concerning for a manufacturing or iatrogenic defect, that likely caused this complication. The patient's health care proxy consented to the publication of this patient's operative course.
No takes yet. Share an insight, caveat, or question.
Bellomo et al. (2025) conducted a case report in Ruptured intra-aortic balloon pump (n=1). Exploratory laparotomy and external iliac arteriotomy was evaluated. Surgical removal via exploratory laparotomy and external iliac arteriotomy was required to retrieve a ruptured intra-aortic balloon pump after unsuccessful endovascular snaring.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: