Key result
Bio-compatible AlN piezoelectric skin sensors show strong agreement with ECG for pulse and breathing rates.
Why the study?
Conventional wearable continuous pulse wave monitoring solutions are typically large and based on technologies that limit their versatility.
Does a biocompatible AlN piezoelectric skin sensor accurately measure cardiovascular physiological parameters compared to standard ECG?
Does a biocompatible AlN piezoelectric skin sensor accurately measure cardiovascular physiological parameters compared to standard ECG?
Effect estimate: r ≥0.81 (pulse rate), r ≥0.95 (breathing rate)
A novel ultrathin, flexible AlN piezoelectric skin sensor accurately monitors pulse rate and breathing rate, showing strong correlation with standard ECG measurements.
Animal data support AlN sensor accuracy; leaves open human validation for cardiovascular monitoring.
Arterial pulse waves contain a wealth of parameters indicative of cardiovascular disease. As such, monitoring them continuously and unobtrusively can provide health professionals with a steady stream of cardiovascular health indices, allowing for the development of efficient, individualized treatments and early cardiovascular disease diagnosis solutions. Blood pulsations in superficial arteries cause skin surface deformations, typically undetectable to the human eye; therefore, Microelectromechanical systems (MEMS) can be used to measure these deformations and thus create unobtrusive pulse wave monitoring devices. Miniaturized ultrathin and flexible Aluminium Nitride (AlN) piezoelectric MEMS are highly sensitive to minute mechanical deformations, making them suitable for detecting the skin deformations caused by cardiac events and consequently providing multiple biomarkers useful for monitoring cardiovascular health and assessing cardiovascular disease risk. Conventional wearable continuous pulse wave monitoring solutions are typically large and based on technologies limiting their versatility. Therefore, we propose the adoption of 29.5 μm-thick biocompatible, skin-conforming devices on piezoelectric AlN to create versatile, multipurpose arterial pulse wave monitoring devices. In our initial trials, the devices are placed over arteries along the wrist (radial artery), neck (carotid artery), and suprasternal notch (on the chest wall and close to the ascending aorta). We also leverage the mechano-acoustic properties of the device to detect heart muscle vibrations corresponding to heart sounds S1 and S2 from the suprasternal notch measurement site. Finally, we characterize the piezoelectric device outputs observed with the cardiac cycle events using synchronized electrocardiogram (ECG) reference signals and provide information on heart rate, breathing rate, and heart sounds. The extracted parameters strongly agree with reference values as illustrated by minimum Pearson correlation coefficients (r) of 0.81 for pulse rate and 0.95 for breathing rate.
No takes yet. Share an insight, caveat, or question.
Shumba et al. (2024) studied Cardiovascular physiology monitoring. Bio-Compatible AlN Piezoelectric Skin Sensors vs. Synchronized electrocardiogram (ECG) was evaluated on Correlation of extracted parameters with reference values (r ≥0.81 (pulse rate), r ≥0.95 (breathing rate)). Bio-compatible AlN piezoelectric skin sensors showed strong agreement with ECG reference values, with minimum correlation coefficients of 0.81 for pulse rate and 0.95 for breathing rate.