Key result
Novel sensor-based tracking feasibly characterizes LV wall motion and correlates strongly with MRI chamber volumes.
Why the study?
Does sensor-based real-time tracking technology accurately characterize left ventricular mechanical and electrical activation patterns compared to echocardiography and MRI in patients with normal systolic function?
Observational (n=10)
Does sensor-based real-time tracking technology accurately characterize left ventricular mechanical and electrical activation patterns compared to echocardiography and MRI in patients with normal systolic function?
Mapping-guided intra-procedural characterization of left ventricular wall motion using sensor-based tracking is feasible and correlates well with standard imaging modalities.
Supports feasibility of sensor-based LV tracking; leaves open larger validation before clinical adoption.
AIMS: In times of evolving cardiac resynchronization therapy, intra-procedural characterization of left ventricular (LV) mechanical activation patterns is desired but technically challenging with currently available technologies. In patients with normal systolic function, we evaluated the feasibility of characterizing LV wall motion using a novel sensor-based, real-time tracking technology. METHODS AND RESULTS: Ten patients underwent simultaneous motion and electrical mapping of the LV endocardium during sinus rhythm using electroanatomical mapping and navigational systems (EnSite™ NavX™ and MediGuide™, SJM). Epicardial motion data were also collected simultaneously at corresponding locations from accessible coronary sinus branches. Displacements at each mapping point and times of electrical and mechanical activation were combined over each of the six standard LV wall segments. Mechanical activation timing was compared with that from electrical activation and preoperative 2D speckle tracking echocardiography (echo). MediGuide-based displacement data were further analysed to estimate LV chamber volumes that were compared with echo and magnetic resonance imaging (MRI). The lateral and septal walls exhibited the largest (12.5 [11.6-15.0] mm) and smallest (10.2 [9.0-11.3] mm) displacement, respectively. Radial displacement was significantly larger endocardially than epicardially (endo: 6.7 [5.0-9.1] mm; epi: 3.8 [2.4-5.6] mm), while longitudinal displacement was significantly larger epicardially (endo: 8.0 [5.0-10.6] mm; epi: 10.3 [7.4-13.8] mm). Most often, the anteroseptal/anterior and lateral walls showed the earliest and latest mechanical activations, respectively. 9/10 patients had concordant or adjacent wall segments of latest mechanical and electrical activation, and 6/10 patients had concordant or adjacent wall segments of latest mechanical activation as measured by MediGuide and echo. MediGuide's LV chamber volumes were significantly correlated with MRI (R2= 0.73, P < 0.01) and echo (R2= 0.75, P < 0.001). CONCLUSION: The feasibility of mapping-guided intra-procedural characterization of LV wall motion was established. CLINICAL TRIAL REGISTRATION: http://www.clinicaltrials.gov; Unique identifier: CT01629160.
No takes yet. Share an insight, caveat, or question.
Piorkowski et al. (2016) conducted an observational in Normal systolic function (n=10). Sensor-based, real-time tracking technology (MediGuide) vs. Echocardiography and MRI was evaluated on Feasibility of characterizing LV wall motion and correlation of LV chamber volumes (R2=0.73 (MRI), R2=0.75 (echo), p=<0.01 (MRI), <0.001 (echo)). Intra-procedural characterization of left ventricular wall motion using a novel sensor-based tracking technology was feasible, with LV chamber volumes significantly correlating with MRI (R2=0.73, P<0.01).