This is the authors' abstract. We don't add key points for this paper.
In a prospective, unblinded, controlled study, 303 children underwent comparison of rectal (RT) and right-sided tympanic membrane (TMT) temperatures. Digital electronic thermometers were used for RT, while two different infrared detection devices assessed TMT. The performance of TMT devices were compared using correlation with RT, fever detection, and ease of use. Correlation between RT and the two devices was good, r = 0.70 and r = 0.69, with little influence of patient age on performance. Using receiver operating characteristics curves to assess fever screening. Thermoscan performed somewhat better than First Temp. Staff rated the Thermoscan as significantly easier to use.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Terndrup et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: