Key result
Higher generator impedance drop and efficacy ratio predict ~44% lower arrhythmia risk post-AF ablation.
Why the study?
Ablation index (AI) and generator impedance drop (ID) predict lesion size, but their correlation with long-term outcomes in persistent AF ablation is not known.
Do lesion monitoring parameters (generator impedance drop and efficacy ratio) predict long-term arrhythmia-free survival after catheter ablation in persistent atrial fibrillation?
Observational (n=427)
Do lesion monitoring parameters (generator impedance drop and efficacy ratio) predict long-term arrhythmia-free survival after catheter ablation in persistent atrial fibrillation?
The efficacy ratio (impedance drop divided by ablation index) is a novel parameter that improves prognostication of arrhythmia recurrence after catheter ablation for persistent atrial fibrillation.
Answers are grounded in this paper and cite their sources.
May aid lesion monitoring in persistent AF ablation; extends observational data but leaves open need for randomized validation.
BACKGROUND: The formation of transmural lesions is necessary for the ablation of persistent atrial fibrillation (prAF). Ablation index (AI) and generator impedance drop (ID) predict lesion size but their correlation with long-term outcomes in prAF is not known. Furthermore, we proposed a new parameter, efficacy ratio (ER) calculated as ID/AI, to gain indirect insight into the role of factors affecting ID but not considered by AI. METHODS: We included ablations performed during the DECAAF II trial if they had uploaded lesion-by-lesion summary data and were performed with radiofrequency catheters on the CARTO system. Average patient-level parameters were calculated from all generated Vizitags. RESULTS: A total of 427 ablations met inclusion criteria and 166 utilized AI. Analyzed as continuous variables, ID and ER predicted long-term arrhythmia-free survival but not AI. The ideal cut-off for ID was ID ≥ 10.4 ohms and had a C-index of 0.55. It predicted reduced risk of arrhythmia: hazard ratio 0.56 [0.36-0.88], p = .013 (arrhythmia-free survival of 67% vs. 52%). Similarly, an ER of 1.7 ohms/100AI had a C-index of 0.58 and predicted reduced arrhythmia recurrence: HR 0.39 [0.22-0.69], p = .001. ER < 1.7 ohms/100AI was related to just 32% arrhythmia-free survival. ER improved prognostication as compared to ID alone and identified a subset of low ID patients with even worse outcomes. CONCLUSION: Average ID was predictive of improved outcomes following ablation of prAF. The ratio of ID/AI (ER) was postulated as a measure to summarize the overall impact of factors not considered in the AI formula and provided improved prognostication.
No takes yet. Share an insight, caveat, or question.
Mekhael et al. (2024) conducted an observational in persistent atrial fibrillation (prAF) (n=427). Generator impedance drop (ID) and efficacy ratio (ER) vs. Lower ID or ER values was evaluated on arrhythmia-free survival (HR 0.56, 95% CI 0.36-0.88, p=0.013). Generator impedance drop ≥10.4 ohms and efficacy ratio ≥1.7 ohms/100AI predicted improved arrhythmia-free survival after ablation for persistent AF.