Key result
Partially wireless pulse oximetry links to fewer loss of signal integrity events versus traditional monitors.
Why the study?
Continuous pediatric pulse oximetry generates hundreds of alarms daily, up to 90% of which are false, often due to motion artifact.
Does partially wireless pulse oximetry reduce episodes of loss of signal integrity compared to traditional wall-connected pulse oximetry in healthy children?
Observational (n=36)
Does partially wireless pulse oximetry reduce episodes of loss of signal integrity compared to traditional wall-connected pulse oximetry in healthy children?
Partially wireless pulse oximetry significantly reduces loss of signal integrity and false alarms due to motion artifact in children compared to traditional wired systems.
Answers are grounded in this paper and cite their sources.
May reduce signal loss events in healthy children; hypothesis-generating and requires randomized trials before practice change.
OBJECTIVES: Continuous pulse oximetry monitoring is routine in many pediatric inpatient units, generating hundreds of alarms per patient per day, up to 90% of which are false (nonactionable). We hypothesized that continuously monitoring pulse oximetry with a partially wireless monitor without physical connection to a wall unit (wireless pox) will decrease episodes of loss of signal integrity (LOSI) generated by motion artifact in healthy pediatric subjects completing age-appropriate activities compared with a traditional wall-connected pulse oximeter (control pox), thereby reducing false alarms. METHODS: Thirty-six healthy children, aged 1 to 17 years, were continuously monitored simultaneously with a wireless pox and a control pox while performing increasing levels of activity for 20 minutes. Continuous pulse oximetry data were recorded in 2 second intervals. Data were analyzed for LOSI. The Wilcoxon signed rank test was then used to compare the control pox to the wireless pox. The control pox had a longer percent time in alarm state related to LOSI in all age groups. CONCLUSIONS: This prospective observational study found that the wireless pox has decreased number and duration of events of LOSI compared with the control pox in healthy children. Implementation of partially wireless pulse oximetry in pediatric inpatient units may decrease episodes of LOSI and false alarms from motion artifact.
No takes yet. Share an insight, caveat, or question.
Petersen et al. (2019) conducted an observational in Healthy children (n=36). Partially wireless pulse oximetry vs. Traditional wall-connected pulse oximeter was evaluated on Loss of signal integrity (LOSI) (p=0.0006). Partially wireless pulse oximetry decreased the number and duration of loss of signal integrity events compared with a traditional wall-connected pulse oximeter in healthy children.
Synapse has enriched 10 related papers. Consider them for comparative context: