We present a candidate Reference Method for determining the concentration of serum creatinine. The method is based on HPLC combined with enzymatic determination. Creatinine plus 14[C]creatinine is extracted by cation-exchange chromatography, subjected to reversed-phase HPLC, and finally quantified enzymatically. Enzymatic measurement ensures no interference from co-eluting compounds, which has been a problem for some reported HPLC assays relying on ultraviolet detection. The average corrected recovery was 100.1% (SEM = 1.1%; n = 15). The accuracy was verified by assaying five sera with target values determined by isotope dilution mass spectrometry. The total interassay CV was less than or equal to 2.5%. We applied the method to study the specificity of HPLC-ultraviolet detection, using 72 plasma samples from hospitalized patients; no interference was noted. Thus, HPLC-ultraviolet detection appears to be specific, provided that sample cleanup is based on cation-exchange chromatography. Our diode-array detector studies of peak homogeneity supported this conclusion. Still, combined HPLC-enzymatic determination ensures even greater accuracy, ranking close to that by isotope dilution mass spectrometry.