Key result
Medtronic Attain Stability Quad lead achieves ~97% implant success with low complications and stable thresholds.
Why the study?
The safety and effectiveness of the novel active fixation quadripolar Medtronic Attain Stability Quad left ventricular lead remained to be determined.
Does the Attain Stability Quad lead provide safe and effective pacing in patients with standard indications for CRT?
Observational (n=440)
Yes
Does the Attain Stability Quad lead provide safe and effective pacing in patients with standard indications for CRT?
The Attain Stability Quad lead demonstrated a high rate of implant success (96.8%) with a low complication rate (2.3%) and very low dislodgement rate (0.7%).
May support Attain Stability Quad lead use in CRT; leaves open need for randomized confirmation.
INTRODUCTION: The Medtronic Attain Stability Quad lead is a quadripolar left ventricular (LV) lead with an active fixation helix assembly designed to fixate the lead within the coronary sinus and pace nonapical regions of the LV. The primary objective of this study was to determine the safety and effectiveness of this novel active fixation quadripolar LV lead. METHODS: Patients with standard indications for cardiac resynchronization therapy (CRT) were enrolled. All patients were followed at 3 and 6 months post-implant and every 6 months thereafter until study closure. Pacing capture thresholds (PCTs) were measured at implant and each follow-up and adverse events (AEs) were recorded upon occurrence. RESULTS: Of the 440 patients who underwent implant procedures, placement of the Attain Stability Quad lead was successful in 426 (96.8%). LV lead-related complications occurred in 10 patients (2.3%), including LV lead dislodgement in three patients (0.7%). The percentage of patients with at least one LV pacing vector with a PCT ≤2.5 V at a 6-month follow-up was 96.3%. The LV lead was successfully fixated to the prespecified pacing location in 97.4% of cases. CONCLUSIONS: This large, multinational study of the Attain Stability Quad lead demonstrated a high rate of implant success with a low complication rate. The active fixation mechanism allowed precise placement of the pacing electrodes at the desired target region with good PCTs and a very low dislodgement rate.
No takes yet. Share an insight, caveat, or question.
Jackson et al. (2020) conducted an observational in Standard indications for cardiac resynchronization therapy (CRT) (n=440). Medtronic Attain Stability Quad lead was evaluated on Safety and effectiveness (implant success). The Medtronic Attain Stability Quad lead achieved a 96.8% implant success rate and a 2.3% complication rate, with 96.3% of patients maintaining acceptable pacing capture thresholds at 6 months.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: