Key result
The mean annual implantable cardioverter-defibrillator malfunction rate was approximately 20-fold higher than the pacemaker malfunction rate (26.5 vs 1.3 per 1000 person-years, P<.001).
Why the study?
What are the annual rates and trends of pacemaker and ICD generator malfunctions?
Meta-Analysis
What are the annual rates and trends of pacemaker and ICD generator malfunctions?
Effect estimate: 20-fold higher
Absolute Event Rate: 26.5% vs 1.3%
p-value: p=<.001
ICD generator malfunction rates are approximately 20-fold higher than pacemaker malfunction rates, highlighting the need for continued monitoring of device performance.
Authors
No takes yet. Share an insight, caveat, or question.
Urges intensified ICD generator surveillance in practice; extends device safety data by quantifying type-specific malfunction disparities.
William H. Maisel (2006) conducted a meta-analysis in Pacemaker and implantable cardioverter-defibrillator (ICD) implants. Implantable cardioverter-defibrillator (ICD) vs. Pacemaker was evaluated on Annual rate of generator malfunction (integral component failure requiring explantation prior to elective replacement) (20-fold higher, p=<.001). The mean annual implantable cardioverter-defibrillator malfunction rate was approximately 20-fold higher than the pacemaker malfunction rate (26.5 vs 1.3 per 1000 person-years, P<.001).