Key result
Ultrahigh resolution mapping linked to lower EGM noise than standard catheters and ~97% AT termination.
Why the study?
Does mapping with the IntellaMap Orion basket and Rhythmia system accurately identify the critical isthmus and facilitate successful ablation in scar-related atrial tachycardia?
Population
33 scar-related atrial tachycardias (post-atrial fibrillation ablation or surgical mitral valve repair)
Comparison
Mapping with the IntellaMap Orion basket and the… vs Standard catheter and surface ECG
Design
Cohort
Authors
Loading...
May aid ablation target identification in scar-related AT; leaves open need for randomized trials before wider adoption.
Observational (n=33)
Does mapping with the IntellaMap Orion basket and Rhythmia system accurately identify the critical isthmus and facilitate successful ablation in scar-related atrial tachycardia?
Absolute Event Rate: 0.011% vs 0.016%
p-value: p=<0.05
Ultrahigh resolution mapping with the Orion basket and Rhythmia system accurately identifies the critical isthmus in scar-related atrial tachycardias, leading to a 97% acute ablation success rate.
Lațcu et al. (2016) conducted an observational in Scar-related atrial tachycardia (n=33). IntellaMap Orion basket and Rhythmia system vs. Standard catheter was evaluated on Noise of bipolar electrogram (mV) (p=<0.05). Ultrahigh resolution mapping with the IntellaMap Orion and Rhythmia system produced lower EGM noise than standard catheters (0.011 vs 0.016 mV, P<0.05) and led to AT termination in 97% of cases.