Key result
Rapid Watchman Flex deployment successfully controls acute bleeding from LAA perforation during angiography.
Why the study?
Pericardial effusion is a common complication of percutaneous LAA closure that often requires interrupting the intervention, highlighting the need to describe acute management strategies.
Case Report (n=1)
Rapid deployment of an LAA closure device can serve as an effective technique for bleeding control following LAA perforation during appendage angiography.
Answers are grounded in this paper and cite their sources.
May offer bailout for LAA perforation; leaves open whether this extends beyond single-case observation.
Background Pericardial effusion is a common complication of percutaneous left atrial appendage (LAA) closure. Acute management is the cornerstone of pericardial effusion treatment and interrupting the intervention is often required. Case summary A 65-year-old man presented an acute 10 mm pericardial effusion following pigtail contrast appendage injection. A rapid Watchman Flex 24 mm (Boston Scientific) deployment permitted bleeding interruption. A needle pericardiocentesis was achieved in order to prevent any haemodynamical instability. Discussion This case report describes an atypical cause of pericardial effusion and a technique for bleeding control with LAA closure device deployment.
No takes yet. Share an insight, caveat, or question.
Champagne et al. (2021) conducted a case report in Pericardial effusion during left atrial appendage angiography (n=1). Watchman Flex 24 mm deployment and needle pericardiocentesis was evaluated. Rapid deployment of a 24 mm Watchman Flex device successfully controlled bleeding from an acute 10 mm pericardial effusion caused by LAA perforation during angiography.