Key result
TauroPace irrigation linked to ~100% lower risk of major CIED infections versus hydrogen peroxide.
Why the study?
The study compared TauroPace, a taurolidine solution for combating cardiac implantable electronic device (CIED) infection, against a historical control of 3% hydrogen peroxide.
Does a taurolidine antimicrobial wash reduce major CIED infection compared to 3% hydrogen peroxide in patients undergoing invasive CIED procedures?
Observational (n=1,163)
No
Does a taurolidine antimicrobial wash reduce major CIED infection compared to 3% hydrogen peroxide in patients undergoing invasive CIED procedures?
A taurolidine-containing antimicrobial wash significantly reduces major CIED infections compared to a hydrogen peroxide wash without increasing adverse events.
TauroPace may lower major CIED infection risk; leaves open confirmation in randomized trials before practice change.
AIMS: TauroPace (Tauropharm, Bavaria Germany), a taurolidine solution for combating cardiac implantable electronic device (CIED) infection, was compared with a historical control of 3% hydrogen peroxide (H2O2) in a prospective observational study. METHODS AND RESULTS: The device pocket was irrigated, and all hardware accessible within (leads, suture sleeves, pulse generator) was wiped with H2O2, TauroPace, or taurolidine in a galenic formulation during any invasive CIED procedure at the study centre. Only CIED procedures covered by TauroPace or H2O2 from 1 January 2017 to 28 February 2022 were included for analysis. Patients who underwent >1 procedure were censored for the last treatment group and reassigned at the next procedure. The primary endpoint was major CIED infection within 3 months. The secondary endpoints were CIED infection beyond 3 months, adverse events potentially related to the antimicrobial solutions, CIED system, procedure, and death, till the end of follow-up. TauroPace covered 654 procedures on 631 patients, and H2O2 covered 551 procedures on 532 patients. The TauroPace group had more patient risk factors for infection than the H2O2 group (P = 0.0058) but similar device and procedure-specific risk factors (P = 0.17). Cardiac implantable electronic device infection occurred in 0/654 (0%) of the TauroPace group and 6/551 (1.1%) of the H2O2 group (P = 0.0075). Death occurred in 23/654 (3.5%) of the TauroPace group and 14/551 (2.5%) of the H2O2 group (P = 0.33). Non-infection related adverse events were rarer in the TauroPace (3.8%) than the H2O2 (6.0%) group (P = 0.0802). CONCLUSION: TauroPace is safe but more effective than H2O2 in reducing CIED infection. CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT05576194.
No takes yet. Share an insight, caveat, or question.
Borov et al. (2023) conducted an observational in Cardiac implantable electronic device (CIED) infection (n=1,163). TauroPace (taurolidine solution) vs. 3% hydrogen peroxide (H2O2) was evaluated on Major CIED infection within 3 months (p=0.0075). TauroPace irrigation during CIED procedures significantly reduced major CIED infections compared to 3% hydrogen peroxide (0% vs 1.1%, P=0.0075).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: