Key result
Micra leadless pacemaker achieves high implant success with a ~3% 30-day major complication rate.
Why the study?
While prior Micra trials demonstrated favorable safety and efficacy, changes in implant populations and safety over time were not well studied.
Does the Micra leadless pacemaker maintain a low complication rate and stable electrical performance in real-world European and Middle Eastern patients?
Population
928 European and Middle Eastern patients receiving a Micra leadless pacemaker
Comparison
MAP EMEA cohort vs Micra IDE and Micra PAR studies
Design
Prospective, single-arm registry
Follow-up
Average of 9.7 ± 6.5 months
Authors
Loading...
Supports Micra safety in real-world EMEA patients; extends pivotal trials but leaves open randomized confirmation.
Observational (n=928)
Yes
Does the Micra leadless pacemaker maintain a low complication rate and stable electrical performance in real-world European and Middle Eastern patients?
In a real-world EMEA registry, the Micra leadless pacemaker showed a 99.9% implant success rate and a low complication rate comparable to prior pivotal trials.
Roberts et al. (2023) conducted an observational in Indication for pacemaker (n=928). Micra leadless pacemaker vs. Historical IDE and PAR cohorts was evaluated on Acute (30-day) major complications. The Micra leadless pacemaker had a 99.9% implant success rate and a 30-day major complication rate of 2.59%, with 12-month complication rates not significantly different from prior IDE and PAR studies.