Answers are grounded in this paper and cite their sources.
The purpose of this study was to measure T 1 and T 2 relaxation times of NAD + proton resonances in the downfield 1 H MRS spectrum in human brain at 7 T in vivo and to assess the propagation of relaxation time uncertainty in NAD + quantification. Downfield spectra from eight healthy volunteers were acquired at multiple echo times to measure T 2 relaxation times, and saturation recovery data were acquired to measure T 1 relaxation times. The downfield acquisition used a spectrally selective 90° sinc pulse for excitation centered at 9.1 ppm with a bandwidth of 2 ppm, followed by a 180° spatially selective Shinnar–Le Roux refocusing pulse for localization. Uncertainty propagation analysis on metabolite quantification was performed analytically and with Monte Carlo simulation. [NAD + ] was quantified in five participants. The mean ± standard deviation of T 1 relaxation times of the H2, H6, and H4 NAD + protons were 205.6 ± 25.7, 211.6 ± 33.5, and 237.3 ± 42.4 ms, respectively. The mean ± standard deviation of T 2 relaxation times of the H2, H6, and H4 protons were 33.6 ± 7.4, 29.1 ± 4.7, and 42.3 ± 11.6 ms, respectively. The relative uncertainty in NAD + concentration due to relaxation time uncertainty was 8.4%–11.4%, and measured brain [NAD + ] ( N = 5) was 0.324 ± 0.050 mM. Using downfield spectrally selective spectroscopy with single‐slice localization, we found T 1 and T 2 relaxation times averaged across all NAD + resonances to be approximately 218 and 35 ms, respectively, in the human brain in vivo at 7 T.
No takes yet. Share an insight, caveat, or question.
Swago et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: