Key result
Sysmex CS-2000i analyzer yields hemostasis reference values consistent with manufacturer recommendations.
Why the study?
Hemostasis reference values may vary depending on the type of analyzer and reagents used, necessitating determination for the Sysmex CS-2000i analyzer.
Cross-Sectional (n=100)
This study establishes reference intervals for routine and specific hemostasis parameters on the Sysmex CS-2000i analyzer, highlighting the need for local validation of reference values.
Local validation of Sysmex CS-2000i hemostasis intervals remains essential; reinforces manufacturer consistency in this single-center cross-sectional cohort.
The article defines reference values for activated partial thromboplastin time, Quick’s value, INR, thrombin time, fibrinogen, antithrombin and II, V, VII, VIII, IX, X, XI and XII coagulation factors, according to existing standards on the automated Sysmex CS‑2000i analyzer. The aim of the study. To determine reference values for routine and specific parameters of the hemostasis, which may vary depending on the type of analyzer and utilized reagents. Materials and methods. After receiving informed consent from donors for medical survey and blood donation, blood samples were obtained from 100 healthy donors: 64 (64%) males и 36 (36%) females. We established reference values with the Sysmex CS‑2000i (Sysmex, Japan) hemostasis analyzer and reagents from Siemens (Siemens Healthcare, Germany). Results. The data obtained were compared with the literature data and the data presented in the instructions for the reagents used. The results obtained for activated partial thromboplastin time (23.59–35.69 sec), fibrinogen (1.67–3.59 g/l) and antithrombin (67.65–114.89%) are comparable to the available data. There are no data on other studied parameters of hemostasis for the Sysmex CS‑2000i analyzer and the reagents used in the work. The obtained reference intervals are consistent with the recommendations of the manufacturer. Conclusions. Reference values vary significantly depending on the analytical systems and reagent kits used, which confirms the need for local derivation or validation of reference intervals for each specific analytical system and in each laboratory.
No takes yet. Share an insight, caveat, or question.
Горгидзе et al. (2023) conducted a cross-sectional in Healthy donors (n=100). Sysmex CS-2000i analyzer vs. Literature data and manufacturer instructions was evaluated on Reference values for routine and specific parameters of hemostasis. The automated Sysmex CS-2000i analyzer yielded reference values for hemostasis parameters (e.g., aPTT 23.59–35.69 sec, fibrinogen 1.67–3.59 g/l) consistent with manufacturer recommendations.