Answers are grounded in this paper and cite their sources.
PURPOSE: To assess the role of magnetic resonance (MR) imaging in detection and quantification of liver iron overload. MATERIALS AND METHODS: MR imaging at 0.5 T was prospectively performed on 77 patients (67 with liver iron overload and 10 without) who underwent a liver biopsy with biochemical determination of the liver iron concentration (LIC) (normal, < 36 mumol per gram of liver tissue [dry weight]). Ratios of signal intensities and liver T2 relaxation time were calculated from images obtained with spin-echo and breath-hold gradient-echo (GRE) sequences. RESULTS: Liver-to-tissue signal intensity ratios were better correlated with LIC than T2 relaxation time. Long-echo-time GRE sequences were the most sensitive for detection of slight overload. Thus, high sensitivity (94%) and specificity (90%) were obtained with a liver-to-fat ratio threshold of 1. The quantification of iron with MR imaging was accurate when the LIC was 80-300 mumol/g. For heavy overload, above 300 mumol/g, quantification was impossible owing to complete signal loss. Pancreatic and splenic signal intensity were unchanged in most cases. CONCLUSION: This method, which can be improved by using more sensitive sequences with a high-field-strength system, should be competitive with biopsy for the diagnosis of substantial liver iron overload.
No takes yet. Share an insight, caveat, or question.
Gandon et al. (1994) studied this question.