Key result
New CIED algorithms for conduction system pacing introduce programming pitfalls like T wave oversensing.
Case Report
Answers are grounded in this paper and cite their sources.
Requires vigilance for T wave oversensing with new CSP configurations; leaves open standardized programming strategies.
Cardiac implantable electronic devices (CIEDs) are indicated for a variety of cardiac conditions.1–3 In the recent years, conduction system pacing (CSP) has been gaining popularity as an alternative to biventricular pacing (BiVP) to provide a more physiological ventricular activation and circumvent the detrimental effects of right ventricular (RV) pacing.2 CSP also appears to be an viable alternative for the “pace and ablate” approach compared to BiVP given comparable outcomes.4 Concurrently, there has been a rise in the number of available CIEDs, CIED algorithms and possible CIED-lead configurations.
No takes yet. Share an insight, caveat, or question.
Chua et al. (2025) conducted a case report in T wave oversensing in dual chamber ICD with left bundle branch pacing. Left bundle branch pacing and ICD programming was evaluated. Conduction system pacing is gaining popularity as an alternative to biventricular pacing, but the rise in available CIED algorithms and lead configurations introduces programming pitfalls like T wave oversensing.