Key result
Openwater wearable optical system shows strong agreement with TCD for monitoring beat-to-beat cerebral blood flow.
Why the study?
Bedside cerebral blood flow monitoring could improve care for acute neurologic diseases, but technical challenges limit current techniques.
Does the Openwater optical system accurately measure cerebral hemodynamics compared to transcranial Doppler in healthy adults?
Observational (n=25)
No
Does the Openwater optical system accurately measure cerebral hemodynamics compared to transcranial Doppler in healthy adults?
Effect estimate: Slope 0.87 (95% CI 0.83-0.92)
The Openwater optical system demonstrates strong correlation with transcranial Doppler for monitoring cerebral hemodynamics in healthy adults, highlighting its potential as a bedside monitor.
Optical monitoring may enable portable cerebral hemodynamics assessment; leaves open clinical utility pending patient validation.
Bedside cerebral blood flow (CBF) monitoring has the potential to inform and improve care for acute neurologic diseases, but technical challenges limit the use of existing techniques in clinical practice. Here we validate the Openwater optical system, a novel wearable headset that uses laser speckle contrast to monitor microvascular hemodynamics. We monitored 25 healthy adults with the Openwater system and concurrent transcranial Doppler (TCD) while performing a breath-hold maneuver to increase CBF. Relative blood flow (rBF) was derived from the changes in speckle contrast, and relative blood volume (rBV) was derived from the changes in speckle average intensity. A strong correlation was observed between beat-to-beat optical rBF and TCD-measured cerebral blood flow velocity (CBFv), R=0.79; the slope of the linear fit indicates good agreement, 0.87 (95% CI:0.83-0.92). Beat-to-beat rBV and CBFv were strongly correlated, R=0.72, but as expected the two variables were not proportional; changes in rBV were smaller than CBFv changes, with linear fit slope of 0.18 (95% CI:0.17-0.19). Further, strong agreement was found between rBF and CBFv waveform morphology and related metrics. This first in vivo validation of the Openwater optical system highlights its potential as a cerebral hemodynamic monitor, but additional validation is needed in disease states.
No takes yet. Share an insight, caveat, or question.
Favilla et al. (2023) conducted an observational in Healthy adults (n=25). Openwater wearable optical system vs. Transcranial Doppler (TCD) ultrasonography was evaluated on Agreement between beat-to-beat optical relative blood flow (rBF) and TCD-measured cerebral blood flow velocity (CBFv) (Slope 0.87, 95% CI 0.83-0.92). The Openwater wearable optical system demonstrated strong agreement with transcranial Doppler for monitoring beat-to-beat relative cerebral blood flow (R=0.79, slope 0.87) during a breath-hold maneuver.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: