Key result
The SENC method yielded a higher average peak circumferential strain magnitude compared to the SPAMM method at 3T.
Cross-Sectional (n=26)
No
Answers are grounded in this paper and cite their sources.
Method-specific strain values should not be used interchangeably at 3T; leaves open validation against outcomes.
We compared peak circumferential strain (Ecc) values with spatial modulation of magnetization (SPAMM) and strain-encoded (SENC) magnetic resonance (MR) imaging at 3 tesla. Correlation coefficients of the averaged peak Ecc values of the 2 methods were statistically significant. However, the average peak Ecc value was significantly lower with SPAMM (-13.5%±3.3%) than with SENC (-21.6%±3.4%) (P<0.0001). The SENC method showed higher circumferential strain than the SPAMM method at 3T.
No takes yet. Share an insight, caveat, or question.
Sugimori et al. (2013) conducted a cross-sectional in Patients undergoing cardiac MR imaging without late gadolinium enhancement (n=26). SENC (strain-encoded) method vs. SPAMM (spatial modulation of magnetization) method was evaluated on Averaged peak circumferential strain (Ecc) (p=<0.0001). The SENC method yielded a higher average peak circumferential strain magnitude compared to the SPAMM method at 3T.