Key result
Hydrogel-coated ablation catheters eliminate steam pops and improve tissue contact vs bare metal ex vivo.
Why the study?
Current radiofrequency ablation catheters are limited by poor contact with trabeculated cardiac tissue, causing uneven heating, collateral damage, and incomplete ablation.
Does a conductive hydrogel coating on radiofrequency ablation catheters improve tissue contact and prevent steam pops compared to bare metal catheters in ex vivo and in vitro preclinical models?
Does a conductive hydrogel coating on radiofrequency ablation catheters improve tissue contact and prevent steam pops compared to bare metal catheters in ex vivo and in vitro preclinical models?
Absolute Event Rate: 0% vs 2.88%
A novel conductive hydrogel coating for radiofrequency ablation catheters improves tissue contact and prevents steam pops in ex vivo models, but requires further optimization of conductivity to match bare metal lesion sizes at higher power.
Radiofrequency ablation is a mainstay of cardiac rhythm management despite high recurrence rates. Current radiofrequency ablation catheters are limited by poor contact with trabeculated cardiac tissue that promotes uneven heating with hot spots that cause collateral damage and regions of incomplete ablation that promote recurrence. Herein, we report on a conductive hydrogel coating of radiofrequency ablation catheters to improve tissue contact while promoting efficient energy transfer. A method was developed to graft a polyether urethane diacrylamide hydrogel to the distal tip of the catheter that maintained stable adhesion following drying, sterilization, and rehydration. The coating also remained intact after passage through an introducer sheath and 50 cycles of radiofrequency ablation at clinical power. The hydrogel-coated catheter demonstrated enhanced tissue contact that was dependent on hydrogel modulus. Hydrogel-mediated ablation prevented steam pop incidence and generated homogeneous lesions in an ex vivo ablation model; however, increased hydrogel conductivity is needed to achieve comparable lesion dimensions as the bare metal catheter and prevent coating damage at higher power. Collectively, these results establish a tunable hydrogel coating method that addresses limitations of conventional radiofrequency ablation and offers a promising approach to enhance the safety and efficacy of cardiac ablation therapies.
No takes yet. Share an insight, caveat, or question.
Bashe et al. (2026) studied Cardiac arrhythmias. Polyether urethane diacrylamide (PEUDAm) conductive hydrogel coating vs. Bare metal ablation catheter was evaluated on Steam pop occurrence and lesion formation in an ex vivo model at 50 W. Hydrogel-coated radiofrequency ablation catheters prevented steam pops (0 vs 2.88 ± 0.47 per 30 seconds) and improved tissue contact compared to bare metal catheters in an ex vivo model.