Key result
Multi-electrode leads detect ~32ms narrower QRS vs standard lumenless leads in a canine model.
Why the study?
LBBA lead placement relies on single-electrode pacing guided by ECG markers during transeptal advancement, prompting the hypothesis that multiple electrodes on a lead tip could facilitate pacing.
Does a multi-electrode lead improve QRS duration and physiological pacing parameters compared to lumenless leads during left bundle branch area pacing in canine hearts?
Does a multi-electrode lead improve QRS duration and physiological pacing parameters compared to lumenless leads during left bundle branch area pacing in canine hearts?
A novel multi-electrode lead facilitated narrower QRS durations during left bundle branch area pacing in a canine model, suggesting potential for improved physiological pacing and streamlined implantation.
Answers are grounded in this paper and cite their sources.
May support multi-electrode leads for LBBAP; leaves open translation to human outcomes.
Abstract Background Left bundle branch area (LBBA) lead placement relies on single-electrode pacing guided by ECG markers during transeptal advancement. We hypothesize that multiple electrodes on a lead tip facilitate LBBA pacing. Objective Develop and evaluate a multi-electrode lead (MEL) with dimensions and electrical properties similar to clinically relevant lumenless leads (LLLs) for use in LBBA pacing. Methods Insulated 125μm silver wires were fixed along a 3D printed MEL shaft. Wires were wrapped circumferentially at distal, medial, and proximal sites (0.4mm apart), secured with polyimide, and de-insulated by sanding. The MEL tip was cut into a bevel shape to facilitate septal insertion. MEL has 5.4mm² surface area across the three electrodes, vs LLL's 3.6mm² surface area. MELs and LLLs were inserted at up to three LBBA sites in perfused canine hearts (n=4), totaling nine sites. Impedance, pacing threshold, and shortest QRS duration at each insertion site were recorded and compared. Results MEL and LLL had similar impedance and capture voltages. MEL's multiple pacing options evoked QRS morphology changes with pacing output and location. MEL more often detected a narrower QRS at insertion than LLL (LLL: 114ms, Q1-Q3: 98-120ms; MEL: 82ms, Q1-Q3: 68-98ms; p=0.03) and recorded QRS durations similar to sinus rhythm, suggesting MEL may better facilitate physiological pacing. Conclusion MELs more often evoked ECG markers indicative of conduction system capture during LBBA pacing, suggesting MELs may streamline implantation, reduce lead repositioning, and advance conduction system pacing strategies.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2025) studied Left bundle branch area pacing (n=4). Multi-electrode lead (MEL) vs. Lumenless leads (LLLs) was evaluated on Shortest QRS duration at insertion site (p=0.03). A multi-electrode lead detected a narrower QRS duration at insertion compared to standard lumenless leads in a canine heart model.
Synapse has enriched 10 related papers. Consider them for comparative context: