This is the authors' abstract. We don't add key points for this paper.
The clinical efficacy of transcranial magnetic stimulation (TMS) critically depends on precise coil positioning; yet, a systematic quantitative analysis of how positioning errors affect the TMS-induced cortical electric field (E-field) intensity remains unavailable. To address this gap, this study developed mathematical models for 28 commercial TMS coils to characterize the impact of six-degree-of-freedom (6-DoF) positioning errors—including translational (X, Y, Z) and rotational (yaw, pitch, roll) components—on cortical E-field intensity. These models were constructed using 55 776 E-field simulations of the primary motor cortex (M1) across 24 healthy adult subjects, enabling the quantification of translational and rotational positioning error sensitivity across coils and the development of a rapid cortical electric field evaluation method (FIEM) for rapid E-field evaluation. Key results demonstrate high accuracy in predicting positioning error effects (R > 0.94 for all 28 coils), significantly broader sensitivity ranges for translational compared to rotational errors, and FIEM achieving 95% accuracy with a 14 ms computational speed. By incorporating cross-vendor, multi-size coil comparisons, our results provide clinicians and researchers with data-driven tools for selecting TMS coils and positioning approaches, while the FIEM framework supports rapid E-field evaluation to reduce uncertainty in cortical targeting during TMS interventions.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Li et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: