Key result
Verily Study Watch shows excellent resting heart rate reliability versus ECG with ~0.95 correlation.
Why the study?
Resting heart rate measured by a wearable device had not been evaluated in a large cohort against diverse baseline characteristics or validated against gold-standard ECG.
Does resting heart rate measured by the Verily Study Watch accurately correlate with ECG measurements and associate with clinical characteristics in adults?
Observational (n=875)
Yes
Does resting heart rate measured by the Verily Study Watch accurately correlate with ECG measurements and associate with clinical characteristics in adults?
Answers are grounded in this paper and cite their sources.
The Verily Study Watch provides highly reliable resting heart rate measurements compared to standard ECG, and higher resting heart rate correlates with adverse clinical characteristics such as higher BMI, CRP, and type 2 diabetes.
BACKGROUND: Though widely used, resting heart rate (RHR), as measured by a wearable device, has not been previously evaluated in a large cohort against a variety of important baseline characteristics. OBJECTIVE: This study aimed to assess the validity of the RHR measured by a wearable device compared against the gold standard of ECG (electrocardiography), and assess the relationships between device-measured RHR and a broad range of clinical characteristics. METHODS: The Project Baseline Health Study (PHBS) captured detailed demographic, occupational, social, lifestyle, and clinical data to generate a deeply phenotyped cohort. We selected an analysis cohort within it, which included participants who had RHR determined by both ECG and the Verily Study Watch (VSW). We examined the correlation between these simultaneous RHR measures and assessed the relationship between VSW RHR and a range of baseline characteristics, including demographic, clinical, laboratory, and functional assessments. RESULTS: From the overall PBHS cohort (N=2502), 875 (35%) participants entered the analysis cohort (mean age 50.9, SD 16.5 years; n=519, 59% female and n=356, 41% male). The mean and SD of VSW RHR was 66.6 (SD 11.2) beats per minute (bpm) for female participants and 64.4 (SD 12.3) bpm for male participants. There was excellent reliability between the two measures of RHR (ECG and VSW) with an intraclass correlation coefficient of 0.946. On univariate analyses, female and male participants had similar baseline characteristics that trended with higher VSW RHR: lack of health care insurance (both P<.05), higher BMI (both P<.001), higher C-reactive protein (both P<.001), presence of type 2 diabetes mellitus (both P<.001) and higher World Health Organization Disability Assessment Schedule (WHODAS) 2.0 score (both P<.001) were associated with higher RHR. On regression analyses, within each domain of baseline characteristics (demographics and socioeconomic status, medical conditions, vitals, physical function, laboratory assessments, and patient-reported outcomes), different characteristics were associated with VSW RHR in female and male participants. CONCLUSIONS: RHR determined by the VSW had an excellent correlation with that determined by ECG. Participants with higher VSW RHR had similar trends in socioeconomic status, medical conditions, vitals, laboratory assessments, physical function, and patient-reported outcomes irrespective of sex. However, within each domain of baseline characteristics, different characteristics were most associated with VSW RHR in female and male participants. TRIAL REGISTRATION: ClinicalTrials.gov NCT03154346; https://clinicaltrials.gov/study/NCT03154346.
No takes yet. Share an insight, caveat, or question.
Feng et al. (2024) conducted an observational in Healthy individuals and those with varying disease risks (n=875). Verily Study Watch vs. Electrocardiography (ECG) was evaluated on Intraclass correlation coefficient between Verily Study Watch resting heart rate and ECG resting heart rate (ICC 0.946). The Verily Study Watch demonstrated excellent reliability for measuring resting heart rate compared to gold-standard electrocardiography, with an intraclass correlation coefficient of 0.946.