Does high-power, short-duration radiofrequency ablation reduce ablation time without compromising safety and efficacy compared to low-power, long-duration ablation in a canine model?
In a canine model, high-power short-duration ablation using a flexible-tip catheter significantly reduced ablation time compared to standard low-power long-duration ablation, without compromising pulmonary vein isolation durability or safety.
Background: High-power, short-duration (HPSD) radiofrequency ablation (RFA) may reduce ablation time. Concerns that catheter-mounted thermocouples (TCs) can underestimate tissue temperature, resulting in elevated risk of steam pop formation, potentially limit widespread adoption of HPSD ablation. Objective: The purpose of this study was to compare the safety and efficacy of HPSD and low-power, long-duration (LPLD) RFA in the context of pulmonary vein isolation (PVI). Methods: , Abbott) was used to isolate the left pulmonary veins (PVs) in 12 canines with HPSD RFA (50 W for 10 seconds) and LPLD RFA (30 W for a maximum of 60 seconds). PVI was assessed at 30 minutes and 28 ± 3 days postablation. Computed tomographic scans were performed to assess PV stenosis after RFA. Lesions were evaluated with histopathology. Results: = .001). All 24 PVs were isolated 30 minutes after ablation, with 12/12 LPLD-ablated and 11/12 HPSD-ablated PVs still isolated at follow-up. Histopathology revealed transmural ablations for HPSD and LPLD RFA. No major adverse events occurred. Conclusion: An investigational ablation catheter effectively delivered RFA lesions. Ablation time required to achieve PVI with HPSD with this catheter was >3-fold shorter than with LPLD RFA.
Ptaszek et al. (Tue,) studied this question.