Key result
Posterior right upper PV interlesion distances of 5.5-6.0 mm correlate with postablation AF recurrence.
Why the study?
An adequate interlesion distance applied during point-by-point pulmonary vein isolation for atrial fibrillation has never been established, and maximum tolerated distance may vary by region.
Does regional maximum tolerated interlesion distance influence postablation atrial fibrillation recurrence in patients with atrial fibrillation undergoing pulmonary vein isolation?
Cohort (n=260)
Does regional maximum tolerated interlesion distance influence postablation atrial fibrillation recurrence in patients with atrial fibrillation undergoing pulmonary vein isolation?
An interlesion distance of 5.5-6.0 mm during pulmonary vein isolation may be inadequate at the posterior aspect of the right upper pulmonary vein, increasing the risk of atrial fibrillation recurrence.
May warrant tighter lesion spacing at the posterior right upper pulmonary vein; hypothesis-generating for ablation optimization and requires prospective validation.
Background: An adequate interlesion distance (ILD) applied during point-by-point pulmonary vein (PV) isolation for atrial fibrillation (AF) has never been established. We hypothesized that maximum tolerated ILD may differ between PV regions and may influence long-term ablation outcomes. Methods: A total of 260 AF patients underwent PV isolation with 3D electroanatomical platform. Postablation, ILD values were classified into 5 groups (6–5.5 mm, 5.5–5.0 mm, 5.0–4.5 mm, 4.5–4.0 mm and <4.0 mm); the number of tags in each group was calculated and correlated with postablation AF recurrence (AFR). All measurements were performed globally for the entire encirclement around each individual PV and regionally for designated PV anatomic segments. Results: Single-procedure freedom from AF was 81% for paroxysmal and 66% for persistent AF at 12 months. Global analysis showed that AFR was not related to median ILD nor the number of lesions within each ILD tag group for any PV. Segmental analysis showed that AFR was not related to median ILD. However, the presence of tags from the 5.5–6.0 mm ILD group located on the posterior aspect of right upper PV (RUPV) correlated with AFR. This was confirmed in a multivariable logistic regression model. Conclusions: Maximum tolerated ILD of 6.0 mm translated into well-accepted ablation results. However, the study suggests that it may be inadequate at the posterior aspect of RUPV.
No takes yet. Share an insight, caveat, or question.
Kiedrowicz et al. (2023) conducted a cohort in Atrial Fibrillation (n=260). Ablation Index Guided Pulmonary Vein Isolation was evaluated on Postablation atrial fibrillation recurrence. Interlesion distances of 5.5-6.0 mm at the posterior right upper pulmonary vein correlated with atrial fibrillation recurrence, despite overall 12-month freedom from AF of 81% in paroxysmal cases.