Why the study?
Radio-frequency ablation of gastric tissue remains in its infancy compared to its extensive cardiac use, prompting investigation into the impact of ablation parameters on gastric lesion formation and tissue damage.
Power-controlled, irrigated radio-frequency ablation safely creates larger and deeper lesions at reduced temperatures in gastric tissue compared to temperature-controlled ablation.
May support irrigated power-controlled RF for larger gastric lesions at lower temperatures; leaves open clinical translation pending human studies.
Radio-frequency ablation of gastric tissue is in its infancy compared to its extensive history and use in the cardiac field. We employed power-controlled, irrigated radio-frequency ablation to create lesions on the serosal surface of the stomach to examine the impact of ablation power, irrigation, temperature, and impedance on lesion formation and tissue damage. A total of 160 lesions were created in vivo in female weaner pigs (n = 5) using a combination of four power levels (10, 15, 20, 30 W) at two irrigation rates (2, 5 mL min−1) and with one temperature-controlled (65 °C) reference setting previously validated for electrophysiological intervention in the stomach. Power and irrigation rate combinations above 15 W resulted in lesions with significantly higher surface area and depth than the temperature-controlled setting. Irrigation resulted in significantly lower temperature (p < 0.001) and impedance (p < 0.001) compared to the temperature-controlled setting. No instances of perforation or tissue pop were recorded for any ablation sequence. Power-controlled, irrigated radio-frequency ablation of gastric tissue is effective in creating larger and deeper lesions at reduced temperatures than previously investigated temperature-controlled radio-frequency ablation, highlighting a substantial improvement. These data define the biophysical impact of ablation parameters in gastric tissue, and they will guide future translation toward clinical application and in silico gastric ablation modeling. Combination of ablation settings (10–30 W power, 2–5 mL min-1 irrigation) were used to create serosal spot lesions. Histological analysis of lesions quantified localized tissue damage.
No takes yet. Share an insight, caveat, or question.
Matthee et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: