Why the study?
Thrombus formation is a risk associated with radiofrequency ablation despite the use of adequate anticoagulants.
Continuous intracardiac echocardiography monitoring during radiofrequency ablation can promptly detect procedure-related thrombus formation, allowing for successful aspiration and potential prevention of embolic complications.
Answers are grounded in this paper and cite their sources.
May support ICE for detecting occult thrombi during ablation; hypothesis-generating, should not yet change practice.
Thrombus formation is a risk associated with radiofrequency ablation (RFA), despite using adequate anticoagulants. We present a case of an 80-year-old woman with persistent atrial tachycardia who underwent catheter ablation, during which continuous intracardiac echocardiography (ICE)-based monitoring of the ablation site revealed an RFA-related thrombus. We approached to the left atrium (LA) via transseptal approach, and ablation catheter (ThermoCool SmartTouch Surround Flow®) was inserted via the Agillis® sheath, which was continuously flushed with heparinized saline. Left atrial activation mapping revealed a mitral-isthmus-dependent flutter; therefore, we planned pulmonary vein and posterior wall (PW) isolation followed by mitral isthmus linear ablation. During linear ablation across the LA PW (30 W; contact force, 10–20 g; and ablation index, 350), ICE (CARTO SOUND®) detected a mobile thrombus at the PW’s centre. The activated clotting time was maintained >350 intraoperatively. The thrombus was successfully aspirated using the Agillis® advanced...
No takes yet. Share an insight, caveat, or question.
Tonegawa‐Kuji et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: