Key result
A pacemaker heart rate score ≥70% is linked to ~270% higher odds of chronotropic incompetence.
Why the study?
Heart rate score (HrSc) ≥70% predicts 5-year mortality in CRT-D and ICD patients, but whether HrSc is related to chronotropic incompetence in pacemaker subjects was unknown.
Does Heart rate score (HrSc) ≥70% correlate with chronotropic incompetence in pacemaker patients?
Observational
Does Heart rate score (HrSc) ≥70% correlate with chronotropic incompetence in pacemaker patients?
Heart rate score ≥70% is an inexpensive, repeatable, and quantitative measure that correlates with treadmill measures of chronotropic incompetence and clinical sick sinus syndrome in pacemaker patients.
HrSc ≥70% may identify chronotropic incompetence in pacemaker patients; leaves open prospective validation for routine use.
BackgroundHeart rate score (HrSc) ≥70% in cardiac resynchronization therapy defibrillator and implantable cardioverter-defibrillator subjects predicts 5-year mortality risk. A high HrSc suggests few sensed cardiac cycles above the programmed lower rate.ObjectiveTo determine if HrSc is related to chronotropic incompetence (CI) in pacemaker (PM) subjects.MethodsHrSc is the percentage of all atrial-paced and sensed events in the single tallest 10 beats/min histogram bin programmed to DDD 60/min. The prospective LIFE study of PM subjects examined multiple treadmill-based measures of CI. The 1-month postimplant DDD 60/min PM rate histogram prior to treadmill was retrospectively analyzed for HrSc. Measures of CI were applied to submaximal treadmill data in the DDD mode. HrSc was compared to these CI measures and to clinical indications for PM.ResultsThe 1-month histogram demonstrated HrSc ≥70% in 43% of subjects. HrSc ≥70% correlated with a clinical diagnosis of sick sinus syndrome (P < .001). CI was present in 34%–88% of subjects by treadmill-based measures. Agreement between treadmill-based measures for CI was poor and varied from 39% to 83%. HrSc ≥70%, as a measure of CI, was most highly correlated with unpaced heart rate <70% of age-predicted maximum heart rate (67%) (odds ratio 3.7, P < .001).ConclusionsHrSc ≥70% correlates with treadmill measures of CI and clinical sick sinus syndrome. HrSc ≥70% is a measure of CI in PM subjects that is inexpensive, repeatable, and quantitative. Heart rate score (HrSc) ≥70% in cardiac resynchronization therapy defibrillator and implantable cardioverter-defibrillator subjects predicts 5-year mortality risk. A high HrSc suggests few sensed cardiac cycles above the programmed lower rate. To determine if HrSc is related to chronotropic incompetence (CI) in pacemaker (PM) subjects. HrSc is the percentage of all atrial-paced and sensed events in the single tallest 10 beats/min histogram bin programmed to DDD 60/min. The prospective LIFE study of PM subjects examined multiple treadmill-based measures of CI. The 1-month postimplant DDD 60/min PM rate histogram prior to treadmill was retrospectively analyzed for HrSc. Measures of CI were applied to submaximal treadmill data in the DDD mode. HrSc was compared to these CI measures and to clinical indications for PM. The 1-month histogram demonstrated HrSc ≥70% in 43% of subjects. HrSc ≥70% correlated with a clinical diagnosis of sick sinus syndrome (P < .001). CI was present in 34%–88% of subjects by treadmill-based measures. Agreement between treadmill-based measures for CI was poor and varied from 39% to 83%. HrSc ≥70%, as a measure of CI, was most highly correlated with unpaced heart rate <70% of age-predicted maximum heart rate (67%) (odds ratio 3.7, P < .001). HrSc ≥70% correlates with treadmill measures of CI and clinical sick sinus syndrome. HrSc ≥70% is a measure of CI in PM subjects that is inexpensive, repeatable, and quantitative.
No takes yet. Share an insight, caveat, or question.
Sharma et al. (2021) conducted an observational in Pacemaker patients. Heart rate score (HrSc) ≥70% vs. Treadmill-based measures of chronotropic incompetence was evaluated on Correlation with unpaced heart rate <70% of age-predicted maximum heart rate (OR 3.7, p=<0.001). A heart rate score ≥70% in pacemaker patients was highly correlated with chronotropic incompetence, specifically an unpaced heart rate <70% of age-predicted maximum heart rate (OR 3.7, P<0.001).