Are current generation coronary guiding catheters suitable for calibration of QCA measurements?
Are current generation coronary guiding catheters suitable for calibration of QCA measurements?
Current generation coronary guiding catheters are suitable for QCA calibration provided it is done on contrast-empty catheters, despite some underestimation of dimensions.
May underestimate QCA dimensions by catheter type; animal data leaves open clinical suitability of current guiding catheters.
Quantitative coronary angiography (QCA) is a validated and widely accepted method to investigate changes in arterial dimension over time. Calibration of measurements is enabled by the use of the coronary catheter as a scaling device. The dimensions and laminar composition of coronary catheters, however, have changed significantly over recent years and the suitability of the current generation of coronary catheters for calibration purposes has not been validated. We therefore recorded 57 coronary guiding catheters on cinefilm, and compared their automated quantitative measurements (Cardiovascular Angiography Analysis System, CAAS) with their true values (precision micrometer). We found an overall underestimation of quantitatively derived dimensions, ranging from -8.9 to +4% for water-filled catheters and from -15.5 to -3.9% for contrast-filled catheters. In conclusion, while the current generation of coronary guiding catheters shows a wide variety in radiological quality, it can be clearly detected by the CAAS system, and is suitable for calibration of QCA measurements (with the exception of the DVI atherectomy catheter), provided that calibration is done on contrast-empty catheters.
No takes yet. Share an insight, caveat, or question.
Herrman et al. (1994) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: