Answers are grounded in this paper and cite their sources.
Two-phase flow widely exists in the process industry, where ultrasonic tomography provides reliable visualization with benefits of noninvasion, nonradiation, and low-cost. To provide real-time reconstruction of low acoustic impedance contrast medium distribution existed in two-phase flow, a novel approach using continuous-wave excitation is proposed. With the basic measurement model introduced, a subspace filtered reconstruction algorithm is introduced with improvements on both accuracy and speed. Through nonuniform shape constrained range filter, the relative image error (RE) is reduced by 32.94% and 33.03% compared to simultaneous algebraic reconstruction technique (SART) and filter back projection (FBP) methods. Through derivation of offline iteration/online reconstruction (OIOR) method and truncated singular value decomposition (TSVD), the computing speed increases to 122 frame/s, which fits the requirement of online monitoring. A prototype of the measurement system is introduced with improved design of bipolar high-voltage excitation module and fan-beam angle piezoelectric transducers. Qualitative and quantitative results show that the proposed method has better image accuracy compared to conventional methods.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: