An ultrasonic human-blood-flow velocity profile measurement method using time-domain correlation of consecutive echo pairs has been developed. The time shift between a pair of range gated echoes is estimated by searching for the shift that results in the maximum correlation. The time shift indicates the distance a group of scatterers has moved, from which flow velocity is estimated. The basis for the computer simulations and error analyses of the scheme includes a band-passed white Gaussian noise signal model for an echo from a scattering medium, the estimate of flow velocity from both a single scatterer and multiple scatterers, and a derived precision estimation. The error analysis via computer simulation includes an evaluation of errors associated with the correlation method. For a uniform flow velocity profile, beamwidth modulation represents the greatest error source. However, for a nonuniform flow velocity profile, the jitter caused by a small flow velocity gradient can exceed the other error sources. A detailed computer simulation evaluated the interdependencies of window length, beam width, vessel diameter, and viewing angle on the estimation of flow velocity.
No takes yet. Share an insight, caveat, or question.
Foster et al. (1990) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: