Experiments measure dose rates for single and composite targets in linear accelerators, revealing key insights on photon production.
The target in electron linear accelerator plays an im-portant role in the production of photon. Targets of different materials and thicknesses have different X-ray yields. In this study, experiments were carried out to measure the dose rates of single targets and composite targets of different thicknesses for 6 MeV linear accel-erators utilizing ionization chamber. The electron ener-gy spectrum at the outlet of accelerating tube was de-tected with magnetic analyser. The experimental results show consistent rules with Monte Carlo simulations. Composite material target of 1.2 mm tungsten and 2 mm copper can deliver 1242 rad/min/100uA dose rate at 1 meter in front of the target. Dose rates of tungsten- rhenium alloy(74%W-26%rhenium) targets were exam-ined too.