Why the study?
Does the amplitude of the atrial potential differ between patients with Sick Sinus Syndrome and Atrio-Ventricular Block during pacemaker implantation?
Does the amplitude of the atrial potential differ between patients with Sick Sinus Syndrome and Atrio-Ventricular Block during pacemaker implantation?
Answers are grounded in this paper and cite their sources.
Atrial potential amplitude is significantly lower in patients with sick sinus syndrome compared to those with atrioventricular block, and correlates with surface ECG P-wave voltages.
SummarySeventy-three measurements of the atrial potential in 64 cases were recorded and studied at the time of physiological pacemaker implantation. The amplitude of the atrial potential in patients with Sick Sinus Syndrome (SSS) was significantly lower than in those with Atrio-Ventricular Block (AVB) (by PSA, 2.2 ± 1.3 mV vs. 3.9 ± 1.1 mV, p < 0.001; by endocardial electrogram, 2.7 ± 1.5 mV vs. 4.6 ± 1.7 mV, p < 0.001). The slew rate was slightly, but not significantly, faster in AVB than in SSS. The amplitude of the atrial potential recorded with bipolar lead was not significantly different from that with unipolar lead (by PSA 2.6 ± 1.5 mV vs. 2.3 ± 1.6 mV; by electrogram, 3.3 ± 1.4 mV vs. 2.8 ± 1.5 mV, NS). In the cases in which P-wave failed to appear after various procedures, the retrograde atrial potential was markedly low (p < 0.001). The atrial potential correlated relatively well with the total of P-wave voltages of lead I, II and III of the body surface ECG (Y = 0.24 ± 0.039 X, r = 0.446, p < 0.01). This can be used in predicting the endocardial atrial potential preoperatively.
No takes yet. Share an insight, caveat, or question.
Murase et al. (1983) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: