Using a two-element phased array receiver coil, single breath-hold, ECG gated cardiac images of signal-to-noise ratios up to 130 and contrast-to-noise ratios exceeding 35 between myocardium and blood were recorded at 3 T. At several locations within the myocardium, T*(2) and B(0) inhomogeneity were determined. Because of shorter T*(2) times and larger B(0) inhomogeneities attributable to enhanced susceptibility effects, real-time cardiac imaging, the use of spiral scans, and echo planar imaging are expected to be considerably more difficult at 3 T.