Key result
Aortic Valve Ejection Fraction (AVEF) is more sensitive to changes in heart contractility than standard Ejection Fraction, yielding a 13.20% increase versus 7.52% for a 10% contractility improvement.
Why the study?
In patients with a continuous flow LVAD, standard EF reflects combined pumping of the heart and device rather than intrinsic heart function, prompting the need for a metric measuring only blood pumped through the aortic valve.
Does the Aortic Valve Ejection Fraction (AVEF) index improve sensitivity to changes in heart contractility compared to standard EF in a mathematical model of an LVAD-assisted left ventricle?
Does the Aortic Valve Ejection Fraction (AVEF) index improve sensitivity to changes in heart contractility compared to standard EF in a mathematical model of an LVAD-assisted left ventricle?
The novel Aortic Valve Ejection Fraction (AVEF) index demonstrates greater sensitivity to changes in heart contractility than standard EF in a mathematical model of an LVAD-assisted left ventricle, suggesting potential utility for non-invasive monitoring.
Ejection Fraction (EF) is an important parameter that is used in the diagnosis of heart failure. It provides a measure of the ability of the heart to pump blood into the circulatory system. For patients implanted with a continuous flow left ventricular assist device (LVAD), the standard definition of EF provides a measure of the combined ability of both the heart to pump blood through the aortic valve and the LVAD to pump blood continuously out of the left ventricle both into the circulatory system. In this paper, we introduce a new concept of EF, labeled Aortic Valve Ejection Fraction (or AVEF) which provides an accurate measure of only the ability of the heart to pump blood through the aortic valve while the LVAD is actively running. We investigate the benefits of AVEF and its relation to the LVAD speed using a well-established mathematical model of an LVAD-assisted left ventricle. We compare the sensitivities of the new AVEF index and the standard EF index to changes in heart contractility over a wide range of after loads. Our results show that AVEF is more sensitive to changes in heart contractility than EF. At a typical LVAD speed of 4,000RPM, a 10% improvement in heart contractility over time for a patient with congestive heart failure yields 13.20% increase in AVEF, while the same increase in contractility yields only 7.52% increase in EF. The AVEF index could provide a reliable, non-invasive mechanism for monitoring improvements in heart contractility for patients implanted with LVAD.
No takes yet. Share an insight, caveat, or question.
Simaan et al. (2019) studied Heart failure with continuous flow left ventricular assist device (LVAD). Aortic Valve Ejection Fraction (AVEF) vs. Standard Ejection Fraction (EF) was evaluated on Increase in index value for a 10% improvement in heart contractility (Emax) at 4,000 RPM. Aortic Valve Ejection Fraction (AVEF) is more sensitive to changes in heart contractility than standard Ejection Fraction, yielding a 13.20% increase versus 7.52% for a 10% contractility improvement.