Key result
A computerized metabolic system accurately measured gas exchange compared to the Douglas bag method, with extremely small differences (e.g., VO2 difference of -0.018 l/min).
Why the study?
Does a computerized metabolic system using inspiratory and expiratory methods accurately measure gas exchange compared to the Douglas bag method in male subjects?
Population
8 male subjects
Comparison
Computerized metabolic system using inspiratory… vs Criterion (Douglas bag, or DB) method
Authors
Loading...
May accelerate lab-based gas exchange testing; leaves open validation in diverse clinical populations.
Cross-Sectional (n=8)
Does a computerized metabolic system using inspiratory and expiratory methods accurately measure gas exchange compared to the Douglas bag method in male subjects?
A computerized metabolic system using inspiratory or expiratory configurations provides highly precise gas exchange measurements comparable to the Douglas bag technique but in less time.
Bassett et al. (2001) reported a cross-sectional. Computerized metabolic system vs. Douglas bag method was evaluated on Gas exchange variables including FE(O2), FE(CO2), and VO2. A computerized metabolic system accurately measured gas exchange compared to the Douglas bag method, with extremely small differences (e.g., VO2 difference of -0.018 l/min).