Plasma high-density lipoprotein is commonly estimated by measuring the cholesterol remaining in plasma supernatant solutions after other lipoproteins, which contain apolipoprotein B, are precipitated with heparin and Mn2+. The method (method I) now in use by the Lipid Research Clinics, in which Mn2+ is at 46 mmol/liter final concentration, is reasonably accurate, but precipitation and sedimentation of lipoproteins other than high-density lipoproteins is often incomplete. We evaluated two modifications of method I. In method II, the Mn2+ concentration was doubled; the second modification (method III) included the increased Mn2+ concentration in a combined heparin Mn2+ reagent, decreased sample volume (2 ml), and a shorter incubation time (10 min at room temperature). The percentages of samples with turbid supernates (i.e., incomplete sedimentation) by methods I, II, and III were 9, 3, and 2%, respectively. Among non-turbid supernates, the percentages of samples containing measurable apolipoprotein B (incomplete precipitation) were 79, 19, and 16%, respectively. We conclude that method III is the most convenient and accurate of the three procedures.