Key result
Ultra-high resolution photon-counting CT reduces deep-learning plaque volume overestimation by ~15% vs standard resolution.
Why the study?
To evaluate the impact of varying spatial resolutions on image quality and deep learning-based quantification of coronary stenosis and plaque volumes using photon-counting detector CT.
Does ultra-high resolution photon-counting detector CT reduce overestimation of coronary stenosis and plaque volumes compared to standard resolution in patients with suspected CAD?
Observational (n=70)
No
Does ultra-high resolution photon-counting detector CT reduce overestimation of coronary stenosis and plaque volumes compared to standard resolution in patients with suspected CAD?
p-value: p=<0.001
Ultra-high resolution photon-counting detector CT significantly reduces the overestimation of deep learning-based quantitative plaque volume and stenosis severity in calcified and partially calcified lesions compared to standard resolution.
May improve deep-learning plaque quantification accuracy; hypothesis-generating and should not change practice without outcome data.
Rationale and Objectives: To evaluate the impact of varying spatial resolutions on image quality, and deep-learning (DL) based quantification of coronary stenosis and plaque volumes using photon-counting detector CT (PCD-CT). Materials and Methods:In this prospective single-center study, 70 patients who were suggestive of coronary artery disease (CAD) and had undergone UHR-CCTA on the PCD-CT system were included.The CCTA datasets were reconstructed with slice thicknesses of standard resolution (SR) 0.6 mm, high spatial resolution (HR) 0.4 mm, and ultra-high resolution (UHR) 0.2 mm.Two radiologists assessed subjective image quality, while a DL-based tool automatically quantified the degree of stenosis and plaque volumes across the three resolutions.The extent of stenosis and plaque volumes were compared between UHR, HR, and SR using the Friedman test with post hoc testing, which had Bonferroni correction (0.05/3 = .017).Results: Subjective assessment revealed that higher spatial resolution enhanced vessel sharpness, and stronger calcified and partially calcified plaque diagnosis confidence, but concurrently increased perceived image noise (both P < 0.05).The DL-based stenosis evaluation of 139 coronary artery plaques revealed a decreasing trend of stenosis diameter with improving spatial resolution for calcified and partially calcified plaques, particularly for calcified plaques.(UHR, HR, and SR, respectively: 15% [IQR, 10-20%], 19% [IQR, 10-20.5%],and 20% [IQR, 11-21%]; P = 0.011), whereas noncalcified plaques did not show evidence of a difference (P = 0.292).The median plaque volume exhibits a downward trend as the resolution increases (for overall plaque volume, 29.05 mm 3 [IQR, 13.42 mm 3 , 58.86 mm 3 ], 31.9 mm 3 [IQR, 16.97 mm 3 , 66.82 mm 3 ],34.21mm 3 [IQR, 17.13 mm 3 , 67.9 mm 3 ], P < 0.001, respectively). Conclusion:Although UHR PCD-CT increases perceived image noise, its superior spatial resolution enhances vessel sharpness and significantly reduces the overestimation of DL-based quantitative plaque volume and stenosis severity in calcified and partially calcified lesions.Recognizing this resolution-dependent variability is essential for reliable CAD risk stratification.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2026) conducted an observational in Coronary artery disease (n=70). Ultra-high resolution (0.2 mm) and high spatial resolution (0.4 mm) CCTA vs. Standard resolution (0.6 mm) CCTA was evaluated on Degree of stenosis and plaque volumes (p=<0.001). Ultra-high resolution (0.2 mm) photon-counting CT significantly reduced deep-learning overestimation of median plaque volume compared to standard resolution (29.05 vs 34.21 mm3; P<0.001).