Key result
Unipolar peak-negative voltage with wide-band filtering showed a 70% reduction after ablation, predicting conduction block with 94.4% sensitivity and 80% specificity (P=0.03 vs other methods).
Why the study?
Does unipolar peak-negative voltage with wide-band filtering improve the detection of acute ablative tissue injury in patients undergoing cavotricuspid isthmus ablation for atrial flutter?
Observational (n=27)
Does unipolar peak-negative voltage with wide-band filtering improve the detection of acute ablative tissue injury in patients undergoing cavotricuspid isthmus ablation for atrial flutter?
p-value: p=0.03
Noncontact unipolar electrogram using peak-negative voltage with wide-band filtering is the most sensitive technique for detecting acute ablative tissue injury during atrial flutter ablation.
May support wide-band unipolar voltage for verifying block after flutter ablation; hypothesis-generating pending prospective validation.
BACKGROUND: The optimal recording technique of the electrogram voltage for detecting abnormal atrial tissue remains unclear. The aim of this study was to compare the impact of various recording techniques on the electrogram voltage after the delivery of ablation therapy in the human right atrium (RA). MATERIAL AND METHODS: Noncontact mapping was performed in 27 patients with typical atrial flutter (mean age = 63 +/- 16, males = 20). Noncontact unipolar and bipolar electrograms were obtained before and after cavotricuspid isthmus (CTI) linear ablation. All unipolar electrograms were acquired with both wide-band filtering (0.5-300 Hz) and narrow-band filtering (32-300 Hz). The unipolar voltage measurements included both the peak-to-peak voltage and peak-negative voltage (PNV) for both filter settings. RESULTS: A comparison of the electrogram voltage along the ablation line before and after the ablation demonstrated a greater reduction in the unipolar PNV with wide-band filtering (70 +/- 24%) than in any of the other recording modalities (P = 0.03). It was the most sensitive and specific recording technique to predict conduction block (cut-off Value 0.35 mV; sensitivity = 94.4% and specificity = 80%). A comparison of the electrogram voltage between the ablated atrial myocardium and nearby nonablated myocardium showed that the unipolar PNV with the wide-band filtering remained the most sensitive method to detect the acute ablative tissue injury, whereas the peak-to-peak bipolar voltage was the most specific method. CONCLUSION: The noncontact unipolar electrogram using the PNV with wide-band filter settings (0.5-300 Hz) provided the most sensitive recording technique for detecting acute ablative tissue injury.
No takes yet. Share an insight, caveat, or question.
Lin et al. (2007) conducted an observational in typical atrial flutter (n=27). Unipolar peak-negative voltage with wide-band filtering (0.5-300 Hz) vs. Other recording modalities (narrow-band filtering, bipolar peak-to-peak voltage) was evaluated on Reduction in electrogram voltage along the ablation line (p=0.03). Unipolar peak-negative voltage with wide-band filtering showed a 70% reduction after ablation, predicting conduction block with 94.4% sensitivity and 80% specificity (P=0.03 vs other methods).