Key result
Contact force strongly predicts cardiac muscle insertion depth via a Mooney-Rivlig model.
Why the study?
The relationship between catheter contact force and the resulting mechanical response of cardiac muscle, specifically insertion depth and surface deformation diameter, remains unclear.
A 3-parameter Mooney-Rivlin hyperelastic model accurately predicts the mechanical deformation of cardiac muscle subjected to catheter contact forces between 10 and 80 g, which has implications for focal energy ablative therapies.
Answers are grounded in this paper and cite their sources.
Supports hyperelastic modeling of catheter contact forces; leaves open translation to clinical ablation lesion prediction.
The cardiac muscle is elastic and deformable. Pushing a catheter in contact with the cardiac muscle surface to conduct focal energy-based ablative therapies, such as RF ablation, requires an adequate electrode-tissue contact surface to transfer the energy to the target site. In this regard, the relationship between the contact force (CF) and the resulting mechanical response is still unclear, in particular, the insertion depth (ID) and the diameter of the surface deformation. The objective of this study was to quantify these relationships using an ex vivo model and a computational model. A rigid bar with a 2.3 mm diameter blunt tip (mimicking a 7Fr standard ablation catheter) was placed at a perpendicular orientation on a fragment of the porcine heart. CF values ranged from 10 to 80g. We used ANSYS to build a Mooney-Rivlig model of 3 parameters based on hyperelastic material and to simulate the same conditions as in the experiments. The experimental results showed a strong linear correlation between CF and insertion depth ID ( <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">R²=0.97, P < 0.001</tex> ), from <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">0.7 ± 0.3</tex> mm at 10 gto <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">6.9 ± 0.1</tex> mm at 80 g. We also found a strong linear correlation between CF and minor and major diameters of the surface deformation assessed, from <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">4.0 ± 0.4</tex> mm at 20 g to <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10.3 ± 0.0</tex> mm at 80 g ( <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">R²=0.96</tex> ), and from <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">6.4 ± 0.7</tex> mm at 20 g to <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">16.7 ± 0.1</tex> mm at 80 g ( <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">R²=0.95</tex> ), respectively. A descent gradient algorithm was used to minimize the mean square error (MSE) between the experimental and computational results of ID for the 10 values of CF. After trying different combinations for the3 parameters of the Mooney-Rivlig model, an optimal fit was achieved after 5 iterations, with an error of less than 0.55 mm for ID. This same mode was then used to predict the diameter of the surface deformation, obtaining an error of less than 0.65 mm. The results confirm that a Mooney-Rivlig model of three parameters based on hyperelastic material predicts the mechanical behavior of cardiac muscle reasonably well when subjected to CFs between 10 and 80 g. This information has important implications in cardiac ablative therapies based on focal energy application using a catheter tip.
No takes yet. Share an insight, caveat, or question.
Ijima et al. (2023) studied Cardiac muscle deformation during ablation. Contact force applied by a 2.3 mm diameter blunt tip was evaluated on Insertion depth and diameter of surface deformation (R2=0.97, p=< 0.001). Applying contact forces between 10 and 80 g to cardiac muscle showed a strong linear correlation with insertion depth (R2=0.97, P<0.001), accurately predicted by a 3-parameter Mooney-Rivlig model.