Key result
Cryoballoon pulmonary vein isolation achieved complete isolation in 91% of procedures, with 52.5% of patients free from arrhythmia-related symptoms at a mean follow-up of 8.9 months.
Why the study?
Does pulmonary vein isolation using a cryoballoon catheter achieve acute procedural success and improve symptoms in patients with paroxysmal and persistent atrial fibrillation?
Observational (n=40)
Does pulmonary vein isolation using a cryoballoon catheter achieve acute procedural success and improve symptoms in patients with paroxysmal and persistent atrial fibrillation?
Cryoballoon pulmonary vein isolation is a feasible technique with high acute success and symptom reduction for atrial fibrillation, though it frequently requires an additional conventional cryocatheter.
Answers are grounded in this paper and cite their sources.
Supports cryoballoon PVI feasibility for AF symptom reduction; leaves open comparative efficacy pending randomized trials.
AIMS: To assess the acute effects, safety, and clinical outcome of atrial fibrillation (AF) ablation using a cryoballoon catheter. METHODS AND RESULTS: Forty patients with paroxysmal or persistent AF underwent pulmonary vein (PV) isolation with a cryoballoon catheter (Arctic Front, CryoCath). Electrocardiograms were recorded in case of symptomatic AF recurrences, and a 24 h Holter recording was performed at last follow-up. Complete PV isolation was achieved in 39 (91%) of the 43 procedures (56% with the cryoballoon catheter alone, 44% with an additional conventional ryocatheter). The number of balloon applications per procedure was 9.6 +/- 1.6. The PV isolation rate was significantly higher (83.9%) if total vessel occlusions were obtained than if intermediate (63.6%, P = 0.01) or poor occlusions were achieved (38.1%, P = 0.0002). The mean procedure time was 239 +/- 48 min. At follow-up (mean 8.9 +/- 4.6 months), 52.5% of patients were free from arrhythmia-related symptoms and another 17.5% had reduction of arrhythmia-related symptoms. Two cases each of phrenic nerve paralysis and dysphagia occurred. CONCLUSIONS: Cryoballoon PV isolation is a feasible technique with a high acute success rate and comparable clinical outcome to radiofrequency ablation. Although complications were rare, the need for an additional conventional cryocatheter warrants further development of the technique.
No takes yet. Share an insight, caveat, or question.
Malmborg et al. (2008) conducted an observational in Symptomatic paroxysmal and persistent atrial fibrillation (n=40). Cryoballoon pulmonary vein isolation was evaluated on Complete pulmonary vein isolation. Cryoballoon pulmonary vein isolation achieved complete isolation in 91% of procedures, with 52.5% of patients free from arrhythmia-related symptoms at a mean follow-up of 8.9 months.
Synapse has enriched 10 related papers. Consider them for comparative context: