Key result
Cryoablation of substrates adjacent to the AV node carried a negligible risk of permanent AV block, with 0 cases of acute permanent or late AV block observed across 1303 procedures.
Why the study?
Does cryoablation of substrates adjacent to the AV node carry a low risk of permanent AV block in patients with atrial arrhythmias?
Cohort (n=1,201)
Does cryoablation of substrates adjacent to the AV node carry a low risk of permanent AV block in patients with atrial arrhythmias?
Cryoablation of substrates adjacent to the AV node is highly safe, carrying a negligible risk of permanent AV block even when transient AV block occurs during the procedure.
Answers are grounded in this paper and cite their sources.
Supports perinodal cryoablation safety in atrial arrhythmias; leaves open randomized confirmation.
AIMS: Radiofrequency (RF) ablation is effective for ablation of atrial arrhythmias. However, RF ablation in the vicinity of the atrioventricular (AV) node is associated with a risk of inadvertent, irreversible high-grade AV block, depending on the type of substrate. Cryoablation is an alternative method. The objective was to investigate the acute and long-term risks of AV block during cryoablation. METHODS AND RESULTS: We studied 1303 consecutive cryoablations of substrates in the vicinity of the AV node in 1201 patients (median age 51 years, range 6-89 years) on acute and long-term impairment to the AV nodal conduction system. The arrhythmias treated were AV nodal reentrant tachycardias (n=1116), paraseptal and superoparaseptal accessory pathways (n=100), and focal atrial tachycardias (n=87). In 158 (12%) procedures, cryomapping (38 cases) or cryoablation (120 cases) were stopped due to transient AV block (first-degree AV block 74 cases, second-degree AV block 67 cases, and third-degree AV block 17 cases) after which another site was tested. Transient AV block occurred within seconds of mapping up to 3 min of ablation. The incidence of AV block was similar for different substrates. In most cases, AV nodal conduction was restored within seconds but in two cases transient AV block lasted 21 and 45 min, respectively. There were no cases of acute permanent AV blocks. No late AV blocks occurred during follow-up (mean 24 months, range 6-96 months). CONCLUSION: Cryoablation adjacent to the AV node carries a negligible risk of permanent AV block. Transient AV block during ablation is a benign finding.
No takes yet. Share an insight, caveat, or question.
Insulander et al. (2013) conducted a cohort in Atrial arrhythmias with substrates adjacent to the AV node (n=1,201). Cryoablation was evaluated on Acute permanent AV block and late AV block. Cryoablation of substrates adjacent to the AV node carried a negligible risk of permanent AV block, with 0 cases of acute permanent or late AV block observed across 1303 procedures.
Synapse has enriched 10 related papers. Consider them for comparative context: