Key result
Miniaturized ECG patch shows strong agreement with gold-standard Biopac for HR and HRV during stress.
Why the study?
Mental stress exacerbates secondary event risk after MI, but whether ECG features from a wearable patch agree with gold-standard devices had not been validated in this population.
Does a miniaturized patch accurately measure ECG-derived heart rate and heart rate variability features compared to a gold-standard Biopac device in post-MI patients and healthy controls during stress?
Observational
Does a miniaturized patch accurately measure ECG-derived heart rate and heart rate variability features compared to a gold-standard Biopac device in post-MI patients and healthy controls during stress?
Effect estimate: r>0.9
p-value: p=<0.001
A novel miniaturized patch accurately measures heart rate and heart rate variability during stress in post-MI patients, showing strong agreement with gold-standard devices.
Patients with prior myocardial infarction (MI) have an increased risk of experiencing a secondary event which is exacerbated by mental stress. Our team has developed a miniaturized patch with the capability to capture electrocardiogram (ECG), seismocardiogram (SCG) and photoplethysmogram (PPG) signals which may provide multimodal information to characterize stress responses within the post-MI population in ambulatory settings. As ECG-derived features have been shown to be informative in assessing the risk of MI, a critical first step is to ensure that the patch ECG features agree with gold-standard devices, such as the Biopac. However, this is yet to be done in this population. We, thus, performed a comparative analysis between ECG-derived features (heart rate (HR) and heart rate variability (HRV)) of the patch and Biopac in the context of stress. Our dataset contained post-MI and healthy control subjects who participated in a public speaking challenge. Regression analyses for patch and Biopac HR and HRV features (RMSSD, pNN50, SD1/SD2, and LF/HF) were all significant (p<0.001) and had strong positive correlations (r>0.9). Additionally, Bland-Altman analyses for most features showed tight limits of agreement: 0.999 bpm (HR), 11.341 ms (RMSSD), 0.07% (pNN50), 0.146 ratio difference (SD1/SD2), 0.750 ratio difference (LF/HF).Clinical relevance- This work demonstrates that ECG-derived features obtained from the patch and Biopac are in agreement, suggesting the clinical utility of the patch in deriving quantitative metrics of physiology during stress in post-MI patients. This has the potential to improve post-MI patients' outcomes, but needs to be further evaluated.
No takes yet. Share an insight, caveat, or question.
Nawar et al. (2023) conducted an observational in Prior myocardial infarction (MI). Miniaturized ECG patch vs. Biopac device was evaluated on Agreement of ECG-derived features (HR and HRV) between patch and Biopac (r>0.9, p=<0.001). A miniaturized ECG patch demonstrated strong agreement with the gold-standard Biopac device for heart rate and heart rate variability features during stress (r>0.9, p<0.001).