Key result
2D-ICE detects lesion formation in ~88% of saline-enhanced RF deliveries and correlates with gross pathology.
Why the study?
With the implementation of saline-enhanced radiofrequency needle-tip ablation, real-time validation of lesion formation was needed for the controllable creation of transmural lesions.
Does 2D-ICE accurately guide and validate SERF ablation lesion formation compared to gross pathology in a canine model?
Does 2D-ICE accurately guide and validate SERF ablation lesion formation compared to gross pathology in a canine model?
2D-ICE can effectively guide and validate saline-enhanced radiofrequency needle-tip ablation lesion formation in real-time, correlating well with gross pathology.
May support real-time 2D-ICE lesion validation in SERF ablation; hypothesis-generating in canines, human trials needed.
AIMS: With the implementation of saline-enhanced radiofrequency (SERF) needle-tip ablation, real-time validation of lesion formation is needed for the controllable creation of transmural lesions. The aim of the study was to analyse the ability of two-dimensional intracardiac echocardiography (2D-ICE) to guide and validate SERF ablation in real-time. METHODS AND RESULTS: Fifty-six SERF energy deliveries at left ventricular sites of 11 dogs guided by 2D-ICE were analysed (power: 15-50 W; time: 25-120 s; irrigation saline: 60°C with 10 mL/min flow rate). Catheter tip/tissue orientation and lesion formation could be well detected by 2D-ICE in 49 (87.5%) energy deliveries. Gross pathology analysis confirmed excellent 2D-ICE lesion localization, the ability to detect transmural lesions (70% sensitivity, 47% specificity) and positive correlation between 2D-ICE and the corresponding gross pathology measurements of 'maximal lesion depth'; (repeated measures correlation: rrm = 0.43, P = 0.012) and 'depth at maximal lesion width' (D@MW; rrm = 0.51, P = 0.003). The median angle between SERF catheter tip and endocardium was 76° [interquartile range (IQR) 58-83°]. The more perpendicular the catheter tip/tissue orientation was the deeper D@MW (rrm = 0.32, P = 0.045). Grade 3 microbubbles on 2D-ICE during ablation, indicating inadequate catheter tip/tissue contact, was associated with smaller lesion volumes than with Grade 1 microbubbles (284.8 mm3 [IQR 151.3-343.1] vs. 2114.1 mm3 [IQR 1437.0-3026.3], P < 0.001). CONCLUSION: With excellent lesion localization and a 70% detection rate of transmural lesions, 2D-ICE is well suited to validate SERF ablation lesion formation in real-time. The catheter tip/tissue angle impacts the lesion formation and through perpendicular catheter positioning, deeper intramural areas of the myocardium can be reached.
No takes yet. Share an insight, caveat, or question.
Dickow et al. (2021) studied this question. 2D-ICE guided saline-enhanced radiofrequency (SERF) needle-tip ablation vs. Gross pathology was evaluated on Detection of catheter tip/tissue orientation and lesion formation by 2D-ICE. Two-dimensional intracardiac echocardiography successfully detected lesion formation in 87.5% of saline-enhanced radiofrequency deliveries and correlated with gross pathology (rrm=0.51, P=0.003).