Key result
RMN-guided CTI ablation achieved complete bidirectional isthmus block in 84% of cases versus 91% with conventional manual ablation (P=0.52), but was associated with prolonged procedure times.
Why the study?
Does remote magnetic catheter navigation improve procedural success or reduce procedure times compared to conventional manual ablation in patients undergoing CTI ablation for atrial flutter?
RCT (n=90)
Does remote magnetic catheter navigation improve procedural success or reduce procedure times compared to conventional manual ablation in patients undergoing CTI ablation for atrial flutter?
RMN-guided CTI ablation reduces radiation exposure but significantly prolongs procedure times and may be less effective long-term compared to conventional manual ablation.
RMN offers no success advantage and extends procedure duration in CTI ablation; challenges its routine use versus manual navigation.
BACKGROUND: Conventional catheter ablation of cavotricuspid isthmus (CTI)-dependent atrial flutter is a widely applied standard therapy. Remote magnetic catheter navigation (RMN) may provide benefits for different ablation procedures, but its efficacy for CTI ablation has not been evaluated in a randomized, controlled trial. METHODS AND RESULTS: Ninety patients undergoing de novo ablation of atrial flutter were randomly assigned to conventional manual (n=45) or RMN-guided (n=45) CTI ablation with an 8-mm-tip catheter. Complete bidirectional isthmus block was achieved in 84% (RMN) and 91% (conventional catheter ablation) of the cases (P=0.52). RMN was associated with shorter fluoroscopy time (median, 10.6 minutes; interquartile range [IQR], 7.6 to 19.9, versus 15.0 minutes; IQR, 11.5 to 23.1; P=0.043) but longer total radiofrequency application (17.1 minutes; IQR, 8.6 to 25, versus 7.5 minutes; IQR, 3.6 to 10.9; P<0.0001), ablation time (55 minutes; IQR, 28 to 76, versus 17 minutes; IQR, 7 to 31; P<0.0001), and procedure duration (114+/-35 versus 77+/-24 minutes, P<0.0001). Procedure duration in the RMN group did not decrease significantly with case experience. Long-term procedure success, defined as achievement of complete CTI block and freedom from atrial flutter recurrence during 6 months of follow-up, was lower in the RMN group (73% versus 89%, P=0.063). Right atrial angiography after ablation revealed no significant differences between groups in terms of right atrial diameter or CTI length, morphology, and angulation. Furthermore, none of these parameters was predictive for difficult (ablation time >20 minutes) or unsuccessful ablation. CONCLUSIONS: RMN-guided CTI ablation is associated with reduced radiation exposure but prolonged ablation and procedure times as compared with conventional catheter navigation. Our findings suggest that ablation lesions produced with an RMN-guided 8-mm catheter are less effective irrespective of CTI anatomy. TRIAL REGISTRATION: clinicaltrials.gov Identifier: NCT00560872.
No takes yet. Share an insight, caveat, or question.
Vollmann et al. (2009) conducted an RCT in Common-Type Atrial Flutter (n=90). Remote magnetic catheter navigation (RMN)-guided CTI ablation vs. Conventional manual CTI ablation was evaluated on Complete bidirectional isthmus block (p=0.52). RMN-guided CTI ablation achieved complete bidirectional isthmus block in 84% of cases versus 91% with conventional manual ablation (P=0.52), but was associated with prolonged procedure times.