Why the study?
Is radiofrequency ablation using a remote magnetic navigation system with an 8-mm-tip magnetic catheter feasible and effective for achieving complete bidirectional isthmus block in patients with right atrial flutter?
Is radiofrequency ablation using a remote magnetic navigation system with an 8-mm-tip magnetic catheter feasible and effective for achieving complete bidirectional isthmus block in patients with right atrial flutter?
Remote magnetic navigation with an 8-mm-tip catheter is feasible, safe, and effective for ablation of atrial flutter, with a short learning curve.
May support feasibility in experienced centers; leaves open confirmation of safety and efficacy in randomized trials.
BACKGROUND: A remote magnetic navigation system (MNS) is available and has been used with a 4-mm-tip magnetic catheter for radiofrequency (RF) ablation of some supraventricular and ventricular arrhythmias; however, it has not been evaluated for the ablation of cavotricuspid isthmus-dependent right atrial flutter (AFL). The present study evaluates the feasibility and efficiency of this system and the newly available 8-mm-tip magnetic catheter to perform RF ablation in patients with AFL. METHODS: Twenty-six consecutive patients (23 men, mean age 64.6 +/- 9.6 years) underwent RF ablation using a remote MNS. RF ablation was performed with an 8-mm-tip magnetic catheter (70 degrees C, maximum power 70 W, 90 seconds). The endpoint of ablation was complete bidirectional isthmus block. To assess a possible learning curve, procedural data were compared between the first 14 (group 1) and the rest (group 2) of the patients. RESULTS: The initial rhythm during ablation was AFL in 20 (19 counterclockwise and 1 clockwise) and sinus rhythm in six patients. Due to technical issues, the ablation in the 18th patient could not be done with the MNS, and so we switched to conventional ablation. The remote magnetic navigation and ablation procedure was successful in 24 of the 25 (96%) remaining patients with AFL. In one patient (patient 2), conventional catheter was used to complete the isthmus block after termination of AFL. The procedure, preparation, ablation, and fluoroscopy times (median [range]) were 53 (30-130) minutes, 28 (10-65) minutes, 25 (12-78) minutes, and 7.5 (3.2-20.8) minutes, respectively. Patients in group 2 had shorter procedure (45 [30-70] min vs 80 [57-130] min, P = 0.0001), preparation (25 [10-30] min vs 42 [30-65] min, P = 0.0001), ablation (20 [12-40] min vs 31 [20-78] min, P = 0.002), and fluoroscopy (7.2 [3.2-12.2] min vs 11.0 [5.4-20.8] min, P = 0.014) times. No complication occurred during the procedure. CONCLUSION: Using a remote MNS and an 8-mm-tip magnetic catheter, ablation of AFL is feasible, safe, and effective. Our data suggest that there is a short learning curve for this procedure.
No takes yet. Share an insight, caveat, or question.
Arya et al. (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: