Why the study?
Patients with ICDs requiring ventricular pacing face a risk of pacing-induced cardiomyopathy from chronic RV pacing, yet integrating conduction system pacing with ICD therapy remains challenging. The feasibility of pairing a DF-1 defibrillation lead with a Medtronic 3830 CSP lead on a dual-chamber ICD generator was uncertain.
Population
17 consecutive patients undergoing DF-1 ICD configuration with a Medtronic 3830 CSP lead at a single center
Comparison
Baseline LVEF <= 50% (n = 9) vs LVEF > 50% (n = 8) subgroups
Design
Single-center retrospective case series
Follow-up
Median 15.0 months (IQR 6.2-19.2)
Answers are grounded in this paper and cite their sources.
Feasibility of DF-1-enabled LBBAP-ICD integration shown in one case; leaves open whether this prevents PICM, warranting prospective trials.
BACKGROUND: Patients with implantable cardioverter-defibrillators (ICDs) who require ventricular pacing are at risk for pacing-induced cardiomyopathy (PICM) from chronic right ventricular (RV) pacing. Conduction system pacing (CSP), particularly left bundle branch area pacing (LBBAP), offers more physiologic ventricular activation, but integration with ICD therapy remains challenging. The DF-1 ICD connector standard provides an opportunity to incorporate a separate CSP lead while maintaining high-voltage ICD therapy. We evaluated the feasibility of an alternative ICD configuration using a DF-1 defibrillation lead paired with a Medtronic 3830 CSP lead connected to a dual-chamber ICD generator. METHODS: We conducted a retrospective case series of 17 consecutive patients who underwent this configuration at a single center between October 2022 and August 2025. For subgroup analyses, patients were stratified by baseline left ventricular ejection fraction (LVEF) as LVEF ≤ 50% (n = 9) or LVEF > 50% (n = 8). QRS duration, LVEF, left ventricular end-systolic dimension (LVESD), tricuspid regurgitation (TR) severity, and device outcomes were assessed during follow-up. RESULTS: Median follow-up was 15.0 months (IQR 6.2-19.2). The Medtronic 3830 lead was successfully implanted in all 17 patients, with confirmed LBBAP in 16 patients (94.1%) and deep septal pacing without definitive left bundle branch capture in one patient. No CSP lead revisions for mechanical or electrical failure were required. Among patients with paired measurements, LVEF increased from 41.6% ± 15.0% to 45.9% ± 14.1%, although the change was not statistically significant (p = 0.169), while LVESD decreased from 4.33 ± 0.72 cm to 3.89 ± 0.79 cm (p = 0.022). In paired subgroup analyses, patients with baseline LVEF ≤50% demonstrated an increase in LVEF from 31.9% ± 6.2% to 40.1% ± 14.3% (p = 0.053) and a decrease in LVESD from 4.62 ± 0.65 cm to 4.11 ± 0.71 cm (p = 0.048), whereas patients with baseline LVEF >50% demonstrated no significant change in either measure. QRS duration did not change significantly in either subgroup. Overall TR severity did not change significantly (Bhapkar p = 0.337), although 5 of 12 patients with paired imaging experienced at least a one-grade increase in TR severity and no patient developed new severe TR. Two patients (11.8%) died during follow-up, with neither death attributed to the device configuration. CONCLUSION: An ICD configuration using a DF-1 defibrillation lead paired with a Medtronic 3830 CSP lead on a dual-chamber generator was technically feasible, with successful 3830 lead implantation in all patients, confirmed LBBAP in 94.1%, and no CSP lead revisions for mechanical or electrical failure during intermediate-term follow-up. Patients with baseline LVEF ≤50% demonstrated numerical improvement in LVEF and a reduction in LVESD, while ventricular function remained stable in those with LVEF >50%. These exploratory findings require confirmation in larger prospective comparative studies.
No takes yet. Share an insight, caveat, or question.
Mrozek et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: