Phase I/II trial evaluates cRGD-ZW800-1 to enhance surgical margins in oral cancer, suggesting improved outcomes.
In oral squamous cell carcinoma (OSCC) surgery, inadequate tumor margins are reported in up to 85% of cases, adversely affecting outcomes. In this prospective, single-center study (n = 31; NCT04191460), we evaluated the safety and imaging feasibility of cRGD-ZW800-1, a near-infrared fluorescent integrin-targeted tracer. The secondary objective was to determine whether intraoperative fluorescence imaging could identify inadequate resection margins and inform surgical decision-making. Patients received 0.01, 0.025, or 0.05 mg/kg cRGD-ZW800-1, and tracer uptake was quantified using in vivo multi-diameter single-fiber reflectance and single-fiber fluorescence spectroscopy to optimize dosing and timing. All doses were well tolerated, achieving tumor-to-background ratios exceeding 4.5, with optimal performance at 0.025 mg/kg. Fluorescence imaging detected all 23 inadequate margins, including nine undetected by conventional assessment (sensitivity 100% versus 70%), altered surgical plans in five cases, and avoided adjuvant radiotherapy in three cases. These findings demonstrate that cRGD-ZW800-1 is safe, tumor-specific, and facilitates intraoperative margin assessment in OSCC.
No takes yet. Share an insight, caveat, or question.
Zweedijk et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: