Key result
A novel V-stent technique using a bent 15 mm Palmaz-Schatz stent mounted on two balloon-catheters was successfully deployed in five cases of porcine coronary bifurcation stenting.
Why the study?
Does the V-stent technique allow successful coronary bifurcation stenting in a porcine model?
Does the V-stent technique allow successful coronary bifurcation stenting in a porcine model?
The V-stent technique demonstrates feasibility for coronary bifurcation stenting in a preclinical porcine model.
Answers are grounded in this paper and cite their sources.
Feasible in porcine bifurcations; leaves open translation to human coronary disease.
Intracoronary stenting as a transcatheter treatment for bifurcation stenoses remains controversial. A novel technique of coronary bifurcation stenting is reported. A 15 mm Palmaz-Schatz stent is bent 180 degrees at its bridge articulation into a V- configuration and mounted on two balloon-catheters linked together by adhesive tape. This unified stent delivery system was used successfully in five cases of porcine coronary bifurcation stenting.
No takes yet. Share an insight, caveat, or question.
Schampaert et al. (1996) studied Coronary bifurcation stenoses (n=5). V-stent (15 mm Palmaz-Schatz stent bent 180 degrees into a V-configuration) was evaluated on Successful deployment. A novel V-stent technique using a bent 15 mm Palmaz-Schatz stent mounted on two balloon-catheters was successfully deployed in five cases of porcine coronary bifurcation stenting.
Synapse has enriched 10 related papers. Consider them for comparative context: